Agilent Technologies 8960 10 Series Reference Manual page 428

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

SETup:PVTime
SETup:PVTime:COUNt:NUMBer
Function
Sets/queries the number of Power vs. Time measurements the test set will make when
multi-measurement state is on.
Setting
Range: 1 to 999
Resolution: 1
Query
Range: 1 to 999
Resolution: 1
*RST Setting
10
Programming Example
OUTPUT 714;"SETUP:PVTIME:COUNT:NUMBER 20" !Sets multi-measurement count value
SETup:PVTime:COUNt:STATe
Function
Sets/queries power versus time multi-measurement count state.
Setting
0|OFF | 1| ON
Query
0|1
*RST Setting
0 |OFF
Programming Example
OUTPUT 714;"SETUP:PVTIME:COUNT:STATE ON" !Sets multi-measurement count state
428
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_pvtime.fm
!to 20.
!to on.