Agilent Technologies 8960 10 Series Reference Manual page 356

Wireless communications test set, gsm mobile test application
Hide thumbs Also See for 8960 10 Series:
Table of Contents

Advertisement

INITiate
INITiate:<measurement mnemonic>[:ON]
Function
Starts measurements with the test set.
The INITiate command is associated with the SETup command, and the FETCh? command, see
"SETup Subsystem" on page 379
One or more measurements may be initiated on the same program line. See
Measurements" on page
This command is also used to activate a measurement. See
(how INIT commands are used)" on page
Programming Example
OUTPUT 714;"INITIATE:TXPOWER:ON" !Initiates a TX Power measurement.
OUTPUT 714;"INITIATE:TXPOWER;PFERROR:ON" !Initiates TX Power and
!phase and frequency error measurements.
OUTPUT 714;"INITIATE:PVTIME;ORFSPECTRUM;FBERROR:ON" !Initiates power
!versus time output RF spectrum, and fast bit error rate measurements.
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
INITiate:<measurement mnemonic>:OFF
Function
Deactivates the selected measurement. See
Only one measurement can be deactivated at a time, to stop one or more measurements and leave
them in the active state, see
Programming Example
OUTPUT 714;"INITIATE:TXPOWER:OFF" !Deactivates TX power measurement.
INITiate:COUNt?
Function
Queries the number of measurements that have been initiated (that are activate). See
"Measurement States" on page
Query
Range: 0 to 10
Resolution: 1
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
356
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_initiate.fm
and
"FETCh? Subsystem" on page
122.
353.
"Measurement States" on page
"ABORt" on page
216.
150.
295.
"Concurrent
"INITiate Programming Examples
150.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5515aE5515bE1960a