Agilent Technologies 85050C User's And Service Manual
Agilent Technologies 85050C User's And Service Manual

Agilent Technologies 85050C User's And Service Manual

7 mm precision calibration kit

Advertisement

Quick Links

User's and Service Guide
Agilent Technologies 85050C
7 mm Precision Calibration Kit
This manual applies directly to 85050C calibration kits with serial number prefix 3106A.
The calibration devices in this kit are individually serialized. Record the device serial
numbers in the table provided in this manual (see "Recording the Device Serial Numbers"
in Chapter 1.)
Agilent Part Number: 85050-90051
Printed in USA
Print Date: October 2006
Supersedes: June 2002
© Copyright 1994, 1996, 2001, 2002, 2006 Agilent Technologies, Inc. All rights reserved.

Advertisement

Table of Contents
loading

Summary of Contents for Agilent Technologies 85050C

  • Page 1 Agilent Technologies 85050C 7 mm Precision Calibration Kit This manual applies directly to 85050C calibration kits with serial number prefix 3106A. The calibration devices in this kit are individually serialized. Record the device serial numbers in the table provided in this manual (see “Recording the Device Serial Numbers”...
  • Page 2 (November 1995), as applicable in any technical data. Certification Agilent Technologies, Inc. certifies that this product met its published specifications at the time of shipment from the factory. Agilent Technologies, Inc. further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology, to the extent allowed by the Institute's calibration facility, and to the calibration facilities of other International Standards Organization members.
  • Page 3 Product maintenance agreements and other customer assistance agreements are available for Agilent products. For any assistance, contact Agilent Technologies. Refer to “Contacting Agilent” on page 4. Printing Copies of Documentation from the Web To print copies of documentation from the Web, download the PDF file from the Agilent web site: •...
  • Page 4 85050C...
  • Page 5: Table Of Contents

    85050C Kits with Serial Prefix 3027A........
  • Page 6 Contents How Often to Recertify ............. . 4-3 Where to Send a Kit for Recertification.
  • Page 7: General Information

    General Information 85050C...
  • Page 8: Calibration Kit Overview

    Calibration Kit Overview Calibration Kit Overview The Agilent 85050C 7 mm calibration kit is used to calibrate Agilent network analyzers up to 18 GHz for measurements of components with 7 mm connectors. The standards in this calibration kit allow you to perform 2-port TRL (thru-reflect-line) calibrations.
  • Page 9: Calibration Definitions

    “Contacting Agilent” on page 5-4. Agilent will arrange for repair or replacement of incomplete or damaged shipments without waiting for a settlement from the transportation company. 85050C...
  • Page 10: Serial Numbers

    Record these serial numbers in Table 1-1. Recording the serial numbers will prevent confusing the devices in this kit with similar devices from other kits. Table 1-1 Serial Number Record for the 85050C Device Serial Number Calibration kit _______________________________...
  • Page 11: Calibration Kits Documented In This Manual

    Calibration Kits Documented in This Manual Calibration Kits Documented in This Manual This manual applies to any 85050C calibration kit whose serial number is listed on the title page. If your calibration kit has a different serial number prefix, refer to the “Calibration Kit History”...
  • Page 12 General Information Calibration Kits Documented in This Manual 85050C...
  • Page 13: Specifications

    Specifications 85050C...
  • Page 14: Environmental Requirements

    During a measurement calibration, the temperature of the calibration devices must be stable and within the range shown in Table 2-1. Avoid unnecessary handling of the devices during calibration because your IMPORTANT fingers are a heat source. 85050C...
  • Page 15: Mechanical Characteristics

    Mechanical characteristics such as center conductor protrusion and pin depth are not performance specifications. They are, however, important supplemental characteristics related to electrical performance. Agilent Technologies verifies the mechanical characteristics of the devices in the kit with special gaging processes and electrical testing.
  • Page 16 If the pin depth of a device does not measure within the observed pin depth limits, it may be an indication that the device fails to meet electrical specifications. Refer to Figure 2-1 for a visual representation of proper pin depth (slightly recessed). 85050C...
  • Page 17: Precision Airline

    60 dB. Do not intermix the center and outer conductors between calibration kits. • Measurement by the user of the diameters is not recommended, and will cause damage to the calibration airline, due to the lack of proper measuring equipment and tooling. 85050C...
  • Page 18 Specifications Mechanical Characteristics Figure 2-2 7 mm Airline Mechanical Dimensions 85050C...
  • Page 19: Electrical Specifications

    Calibration that came with your kit for the exact dimensions of your airline. Certification Agilent Technologies certifies that this product met its published specifications at the time of shipment from the factory. Agilent further certifies that its calibration measurements are traceable to the United States National Institute of Standards and Technology (NIST) to the extent allowed by the institute’s calibration facility, and to the calibration facilities...
  • Page 20 Specifications Electrical Specifications 85050C...
  • Page 21: Use, Maintenance, And Care Of The Devices

    Use, Maintenance, and Care of the Devices 85050C...
  • Page 22: Electrostatic Discharge

    4. Remove the short from the cable. Figure 3-1 shows a typical ESD protection setup using a grounded mat and wrist strap. Refer to Chapter 6, “Replaceable Parts,” for information on ordering supplies for ESD protection. Figure 3-1 ESD Protection Setup 85050C...
  • Page 23: Visual Inspection

    If a connector shows deep scratches or dents, particles clinging to the mating plane surfaces, or uneven wear, clean and inspect it again. Devices with damaged connectors should be discarded. Determine the cause of damage before connecting a new, undamaged connector in the same configuration. 85050C...
  • Page 24: Cleaning Connectors

    Apply a small amount of isopropyl alcohol to a lint-free cleaning swab. b. Clean the connector threads. c. Let the alcohol evaporate, then blow the threads dry with a gentle stream of clean, low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it. 85050C...
  • Page 25 Let the alcohol evaporate, then blow the connector dry with a gentle stream of clean, low-pressure compressed air or nitrogen. Always completely dry a connector before you reassemble or use it. 4. Inspect Inspect the connector to make sure that no particles or residue remain. Refer to “Visual Inspection” on page 3-3. 85050C...
  • Page 26: Gaging Connectors

    Use, Maintenance, and Care of the Devices Gaging Connectors Gaging Connectors The gages available from Agilent Technologies are intended for preventive maintenance and troubleshooting purposes only. See Table 6-1 on page 6-2 for part number information. They are effective in detecting excessive center conductor protrusion or recession, and conductor damage on DUTs, test accessories, and the calibration kit devices.
  • Page 27: Reading The Connector Gage

    If the center conductor protrudes, the indicator will move clockwise to indicate the amount of protrusion, which is read as a positive value. Refer to “Pin Depth” on page 2-3 for definitions of protrusion and recession. Figure 3-2 Reading the Connector Gage 85050C...
  • Page 28: Gaging Procedure

    Using an open-end wrench to keep the gage from rotating, use the torque wrench recommended for use with the kit to tighten the device connector nut to the specified torque. Refer to “Final Connection Using a Torque Wrench” on page 3-13 for more information. 85050C...
  • Page 29 Compare the average reading with the pin depth specifications listed in Table 2-2 on page 2-5. g. Remove the device from the gage and replace the center conductor collet. 85050C...
  • Page 30 Use, Maintenance, and Care of the Devices Gaging Connectors Figure 3-3 Gaging 7 mm Connectors 3-10 85050C...
  • Page 31: Gaging The Airline

    0.0001 inch for your 0.00005 inch for the mechanical tolerances of the airline (≤2.5 ± gage accuracy, and minor divisions). Reverse the above procedure to remove the gage and short circuit from the airline. Figure 3-4 Gaging the Short Circuit 85050C 3-11...
  • Page 32 Use, Maintenance, and Care of the Devices Gaging Connectors Figure 3-5 Gaging the Airline 3-12 85050C...
  • Page 33: Connections

    Table 3-1 provides information about the torque wrench recommended for use with the calibration kit. A torque wrench is included in the calibration kit. Refer to Table 6-1 on page 6-2 for replacement part number and ordering information. 85050C 3-13...
  • Page 34 3-6. Wrenches opposing each other (greater than 90 degrees apart) will cause a lifting action which can misalign and stress the connections of the devices involved. This is especially true when several devices are connected together. Figure 3-6 Wrench Positions 3-14 85050C...
  • Page 35: How To Separate A Connection

    1. Use an open-end wrench to prevent the device body from turning. 2. Use another open-end wrench to loosen the connecting nut. 3. Complete the separation by hand, turning only the connecting nut. 4. Pull the connectors straight apart without twisting, rocking, or bending either of the connectors. 85050C 3-15...
  • Page 36: Using The Trl Adapter

    Always use good electrostatic discharge (ESD) practices. See “Electrostatic Discharge” on page 3-2. 1. Be sure the test port bias voltage is OFF. 2. Note the orientation of the TRL adapter in Figure 3-8. Figure 3-8 TRL Adapter 3-16 85050C...
  • Page 37 3. As shown in Figure 3-9, connect side “B” of the TRL adapter to the end of the cable on port 2. Side “A” of the TRL adapter is now port 2. Figure 3-9 Connecting the TRL Adapter 85050C 3-17...
  • Page 38 7. Connect side “A” of the airline to test port 1 (or to the end of the cable attached to the port). Torque both connections to 136 N-cm (12 in-lb). Reverse the above procedure to disconnect the airline from the cables. 3-18 85050C...
  • Page 39: Pin/Collet Assemblies

    Using the Centering Beads The centering beads supplied are used when gaging the airlines contained in the 85051B verification kit. Do not use the centering beads with any device in this 85050C 7 mm precision calibration kit. 85050C 3-19...
  • Page 40: Handling And Storage

    • Do not set connectors contact-end down on a hard surface. The plating and the mating plane surfaces can be damaged if the interface comes in contact with any hard surface. 3-20 85050C...
  • Page 41: Performance Verification

    Performance Verification 85050C...
  • Page 42: Introduction

    These two steps establish a traceable link to NIST for Agilent to the extent allowed by the institute’s calibration facility. The specifications data provided for the devices in the kit is traceable to NIST through Agilent Technologies. 85050C...
  • Page 43: Recertification

    A list of NIST traceable numbers may be purchased upon request to be included in the calibration report. Agilent Technologies offers a Standard calibration for the recertification of the kit. For more information, contact Agilent Technologies. Refer to “Contacting Agilent” on page 5-4 for a list of offices.
  • Page 44 Performance Verification Recertification 85050C...
  • Page 45: Troubleshooting

    Troubleshooting 85050C...
  • Page 46: Troubleshooting Process

    Troubleshooting Troubleshooting Process Troubleshooting Process If you suspect a bad calibration, or if your network analyzer does not pass performance verification, follow the steps in Figure 5-1. Figure 5-1 Troubleshooting Flowchart 85050C...
  • Page 47: Returning A Kit Or Device To Agilent

    Troubleshooting Returning a Kit or Device to Agilent Returning a Kit or Device to Agilent If your kit or device requires service, contact Agilent Technologies for information on where to send it. See page 5-4 for contact information. Include a service tag (located near the end of this manual) on which you provide the following information: •...
  • Page 48: Contacting Agilent

    If you do not have access to the Internet, please contact your Agilent field engineer. NOTE In any correspondence or telephone conversation, refer to the Agilent product by its model number and full serial number. With this information, the Agilent representative can determine whether your product is still within its warranty period. 85050C...
  • Page 49: Replaceable Parts

    Replaceable Parts 85050C...
  • Page 50: Introduction

    Introduction Introduction Table 6-1 lists the replacement part numbers for the 85050C calibration kit. Table 6-2 lists the replacement part numbers for items not included in the calibration kit that are either required or recommended for successful operation of the kit.
  • Page 51 30 ml 8500-5344 Foam-tipped cleaning swabs 9301-1243 Grounding wrist strap 9300-1367 5 ft grounding cord for wrist strap 9300-0980 2 ft by 4 ft conductive table mat with 15 ft grounding wire 9300-0797 ESD heel strap (for conductive floors) 9300-1126 85050C...
  • Page 52 Replaceable Parts Introduction Figure 6-1 Replaceable Parts for the 85050C Calibration Kit 85050C...
  • Page 53: Standard Definitions

    Standard Definitions 85050C...
  • Page 54: Version Changes

    Version Changes Version Changes Class assignments and standard definitions may change as more accurate model and calibration methods are developed. The disk shipped with the kit or the latest firmware for your network analyzer will contain the most recent version. 85050C...
  • Page 55: Standard Class Assignments

    Adapter Adapter TRL Option Cal Z _____ System Z __X__ Line Z Set ref: __X__ Thru _____ Reflect Lowband frequency: _____ a. The forward isolation standard is also used for the isolation part of the response and isolation calibration. 85050C...
  • Page 56 Standard Class Assignments for the PNA Series Network Analyzer SOLT Class TRL Class Assignments Assignments Calibration Kit Name: Calibration Kit Name: 7 mm Model 85050C 7 mm Model 85050C Class Class TRL thru TRL reflect 1, 2, 7 2, 6...
  • Page 57: Blank Forms

    _____ Line Z Set ref: _____ Thru _____ Reflect Lowband frequency ___________ a. The forward isolation standard is also used for the isolation part of the response and isolation calibration. b. Broadband loads are used for frequencies up to 2 GHz. 85050C...
  • Page 58 You can select these same class assignments in the PNA calibration menu titled “Modify SOLT Calibration Class Assignments.” Select the checkbox titled “Link FWD TRANS, FWD MATCH, REV TRANS and REV MATCH.” 85050C...
  • Page 59: Nominal Standard Definitions

    The values in the standard definitions table are valid only over the specified operating temperature range. Setting the System Impedance This kit contains only 50 ohm devices. Ensure the system impedance (Z ) is set to 50 ohms. Refer to your network analyzer’s user’s guide for instructions on setting system impedance. 85050C...
  • Page 60 Load or arbitrary impedance only. d. For waveguide, the lower frequency is the same as F e. Typical values only. Disk values may be different. f. Consists of standards number 9 and 15 combined. g. Consists of standards number 1 and 15 combined. 85050C...
  • Page 61 Standard Definitions Nominal Standard Definitions Table A-6 Standard Definitions for the PNA Series Network Analyzer Calibration Kit Label: 7 mm Model 85050C = 50.0 Ω System Z Frequency Offset Standard in GHz Broadband Fixed Coax Broadband load load Load Fixed 2.000...
  • Page 62: Blank Forms

    GHz a. Ensure system Z of network analyzer is set to this value. b. Open, short, load, delay/thru, or arbitrary impedance. c. Load or arbitrary impedance only. d. For waveguide, the lower frequency is the same as F A-10 85050C...
  • Page 63 Calibration Kit Label: ________________________ Frequency Offset Standard in GHz a. Ensure system Z of network analyzer is set to this value. b. Open, short, load, delay/thru, or arbitrary impedance. c. For waveguide, the lower frequency is the same as F 85050C A-11...
  • Page 64 Standard Definitions Nominal Standard Definitions A-12 85050C...
  • Page 65 Agilent disk, part number contacting Agilent Technologies entering contents permanently stored calibration kit calibration definitions drawing of calibration kit incomplete contents what to do drawing of...
  • Page 66 Index data male connectors recertification to determine pin depth defective connectors general information definitions standard deviation from nominal phase handling 3-20 device humidity specifications cleaning connecting 3-13 damage impedance disconnecting 3-15 system handling 3-20 setting maintenance incoming inspection performance information, troubleshooting verifying inspection specifications...
  • Page 67 Index locations services included National Institute of Standards and Technology recession (NIST) center conductor nitrogen, for cleaning pin depth nominal standard definitions numbers regulations replaceable parts environmental serial replaceable parts drawing of report, calibration requirements observed limits environmental pin depth return kit or device to Agilent offset opens and shorts return loss, specifications...
  • Page 68 Index discharge electricity zeroing storage 3-20 connector gage supplies cleaning swabs cleaning system impedance service temperature affect on electrical performance calibration cautions about changes in device error-corrected measurement specifications operating storage verification and measurement test data threads connector cleaning inspecting torque wrench specifications 3-13...

Table of Contents