OUTPut[:STATe]:DELay:RISE <delay>|MIN|MAX, (@<chanlist>)
OUTPut[:STATe]:DELay:RISE? [MIN|MAX,] (@<chanlist>)
NOTE
OUTPut:INHibit:MODE LATChing|LIVE|OFF
OUTPut:INHibit:MODE?
Series N6700 User's Guide
However, when outputs of different module types are sequenced using this
command, there may be an additional offset of a few milliseconds from one
output to another. This offset is the same for each module type and is
repeatable. Once you have characterized this offset, using an oscilloscope for
example, you can adjust the programmed delays to compensate for the offset
and give the desired output sequencing.
Outputs within the same module type can also have an offset if one model has
output relays (Option 761) and another does not. These offsets are also
repeatable and can be compensated for by adjusting the programmed delay
values.
This command sets the delay in seconds that the instrument waits
before enabling the specified output. It affects all off-to-on transitions
including changes in the OUTPut:STATe as well as transitions due to
OUTPut:PROTection:CLEar. Delay times can range from 0 to 1.023
seconds in increments of 1 millisecond.
This command allows multiple output channels to turn on in
sequence. Each output will not turn on until its delay time has
elapsed.
The *RST value = 0.
Refer to the note under OUTPut:DELay:FALL, which also applies to
OUTPut:DELay:RISE.
This command selects the mode of operation of the Inhibit input
(INH). The inhibit function shuts down ALL output channels in
response to an external signal on the Inhibit input. If an output
channel has been turned off by OUTPut:STATe, the inhibit function
does not affect the output channel while it is in the OFF state. The
Inhibit mode setting is stored in non-volatile memory.
The following modes can be selected:
LATChing Causes a logic-true transition on the Inhibit input to disable all outputs. The
outputs remain disabled until the Inhibit input is returned to logic-false and the
latched INH status bit is cleared by sending the OUTP:PROT:CLE command or a
protection clear command from the front panel.
LIVE Allows the enabled outputs to follow the state of the Inhibit input. When the
Inhibit input is true, the outputs are disabled. When the Inhibit input is false,
the outputs are re-enabled.
OFF The Inhibit input is ignored.