Appendix D Service
178
From Sheet 1
Single output
YES
failure?
NO
Output is
YES
unregulated?
NO
Not getting
YES
full output?
NO
Spec failure
YES
on output?
NO
Can't calibrate
YES
one output?
NO
Can't calibrate
YES
any outputs?
NO
OV protection
YES
falsely tripping?
Yes
Go to Sheet 3
Overall Troubleshooting Sheet 2
Perform verification procedures in User's guide.
Calibrate the output if required.
Replace the output module.
Perform verification procedures in User's guide.
Calibrate the output if required.
Replace the output module.
Perform verification procedures in User's guide.
Calibrate the output if required.
Replace the output module.
Perform verification procedures in User's guide.
Calibrate the output if required.
Observe all directions for + and - sense connections.
Replace the output module.
Check for latest firmware revision.
Update firmware if required.
Replace the output module.
Check calibration jumper settings.
Reset calibration password using calibration jumpers
Update firmware if required.
Replace carrier assembly on N6700A mainframes.
Replace interface assembly on N6700B mainframes
Check OV protection setting.
Replace the output module.