Front Panel Keys
NOTE
NOTE
Series N6700 User's Guide
Lockout
The ability to lock the front panel from the front panel is only available on
Agilent N6700 MPS mainframes with firmware revision B.00.00 and up.
You can lock the front panel keys to prevent unwanted control of the
instrument from the front panel. This is the most secure way of
locking the front panel keys because you need a password to unlock
the front panel. The lockout setting is saved in non-volatile memory
so that the front panel remains locked even after AC power is cycled.
If the password is lost, a SCPI command is available to reset the front
panel lockout password. Refer to Chapter 6 under
"SYSTem:PASSword:FPANel:RESet" for more information.
Front Panel:
Select System\Preferences\Lock
In the dialog box, enter the password that
will be required to unlock the front panel.
Then select Lock.
The menu to unlock the front panel
appears every time a key is pressed. Enter
the password to unlock the front panel.
SCPI commands can also lock and unlock the front panel. The SCPI commands
are completely independent of the front panel lockout function. If you use SCPI
commands to lock the front panel, the front panel will be unlocked when AC
power is cycled. Refer to chapter 6 under "SYST:COMM:RLST RWLock".
Keys
You can enable or disable the front panel key clicks.
Front Panel:
Select System\Preferences\Keys
Check Enable key clicks to enable key
clicks. Uncheck to disable key clicks.
You can configure the On/Off key to enable or disable ALL outputs.
Front Panel:
Select System\Preferences\Keys.
Check On/Off key affects all channels.
The ON/Off key will now be active on ALL
output channels.
Operating the Power System 4
SCPI Command:
Not Available
SCPI Command:
Not Available
SCPI Command:
Not Available
55