Series N6700 User's Guide
Step 6. Measure the output voltage with the DVM and enter the data.
Front Panel:
Select the Measured Data field. Enter
the data from the external DVM.
Press Select when done.
Press Next to finish calibration.
Low Range Voltage Programming Calibration
This only applies to Agilent Models N6761A and N6762A.
Step 7. Select the low-voltage programming range. This example selects the 5
volt programming range of channel 1. The value to program a range
must be the maximum voltage of the range.
Front Panel:
Select
System\Admin\Cal\Function\Vprog.
Then select the Low range.
Step 8. Select the first voltage calibration point.
Front Panel:
Select Next. The information field
should indicate: Enter P1 data
Step 9. Measure the output voltage with the DVM and enter the data.
Front Panel:
Select the Measured Data field. Enter
the data from the external DVM.
Press Select when done.
Step 10. Select the second voltage calibration point.
Front Panel:
Select Next. The information field
should indicate: Enter P2 data
Step 11. Measure the output voltage with the DVM and enter the data.
Front Panel:
Select the Measured Data field. Enter
the data from the external DVM.
Press Select when done.
Press Next to finish calibration.
Verification and Calibration Appendix B
SCPI Command:
CAL:DATA <data>
SCPI Command:
CAL:VOLT 5
, (@1)
SCPI Command:
CAL:LEV P1
*OPC?
SCPI Command:
CAL:DATA <data>
SCPI Command:
CAL:LEV P2
*OPC?
SCPI Command:
CAL:DATA <data>
153