CALibrate:LEVel P1|P2|P3
NOTE
CALibrate:PASSword <password>
CALibrate:SAVE
CALibrate:STATe ON|OFF [,<password>]
CALibrate:STATe?
NOTE
CALibrate:VOLTage[:LEVel] <value>, (@<channel>)
CALibrate:VOLTage:CMRR (@<channel>)
Series N6700 User's Guide
This command is used to advance to the next level in the calibration.
P1 is the first calibration level; P2 is the second level; P3 is the third
level.
Some calibration sequences may require some settling time after sending
CAL:LEV but before reading the data from the DVM and sending CAL:DATA.
This command lets you change the calibration password. A new
password is automatically stored in nonvolatile memory and does not
have to be stored with CALibrate:SAVE. If the password is set to 0,
password protection is removed and the ability to enter calibration
mode is unrestricted. The factory-default password 0 (zero).
This command saves calibration constants in non-volatile memory
after the calibration procedure has been completed. If calibration
mode is exited by programming CALibration:STATe OFF without first
saving the new constants, the previous constants are restored.
This command enables or disables calibration mode. Calibration
mode must be enabled for the power system to accept any calibration
commands. The first parameter specifies the ON (1) or OFF (0) state.
The second parameter is the password.
A numeric password is required if calibration mode is being enabled
and the existing password is not 0. If the password is not entered or
is incorrect, an error is generated and the calibration mode remains
disabled. The query returns only the state, not the password.
The *RST value = OFF.
When the calibration state is changed from enabled to disabled, new calibration
constants are lost unless they have already been stored with CALibrate:SAVE.
This command initiates calibration of the output voltage. The value
that you enter selects the range that is being calibrated.
This command initiates calibration of the voltage common mode
rejection ratio.
Language Dictionary 6