Test Record Form – Agilent N6736B and N6746B
Agilent N6736B and N6746B
Description
Voltage Programming & Readback
Minimum Voltage Vout
Front Panel Display Readback
High Voltage Vout
Front Panel Display Readback
CV Load Effect
CV Source Effect
CV Ripple and Noise
peak-to-peak
rms
Transient Response
Voltage at 200 µs
Current Programming & Readback
Minimum Current Iout
Front Panel Display Readback
High Current Iout
Front Panel Display Readback
CC Load Effect
CC Source Effect
Test Description
Voltage Programming & Readback, Zero Voltage
Voltage Programming & Readback, High Voltage
CV Load Effect, Source Effect, Ripple and Noise
Transient Response
Current Programming & Readback, Zero Current
Current Programming & Readback, High Current
CC Load Effect, Source Effect
150
Report No _______________
Model
Minimum Specs.
Constant Voltage Tests
Both
0 mV
Both
Vout − 100 mV
Both
99.8 V
Both
Vout − 200 mV
N6736B
− 20 mV
N6746B
− 30 mV
Both
− 10 mV
Both
N/A
Both
N/A
N6736B
− 500 mV
N6746B
− 1000 mV
Constant Current Tests
Both
− 8.5 mA
Both
Iout − 2 mA
N6736B
0.4893 A
N6746B
0.9885 A
N6736B
Iout − 2.75 mA
N6746B
Iout − 3.5 mA
Both
− 2 mA
Both
− 1 mA
N6736A
100 mV, 0.5 A
100 V, 0.5 A
100 V, 0.5 A
100 V, from 0.25 A to 0.5 A
1.5 mA, 100 V
0.5 A, 100 V
0.5 A, 100 V
Date __________________
Results
Maximum Specs.
__________
+ 200 mV
__________
Vout + 100 mV
__________
100.2 V
__________
Vout + 200 mV
__________
+ 20 mV
__________
+ 30 mV
__________
+ 10 mV
__________
+ 30 mV
__________
+ 18 mV
__________
+ 500 mV
__________
+ 1000 mV
__________
+ 11.5 mA
__________
Iout + 2 mA
__________
0.5107 A
__________
1.0115 A
__________
Iout + 2.75 mA
__________
Iout + 3.5 mA
__________
+ 2 mA
__________
+ 1 mA
Instrument Settings
N6746A
100 mV, 1 A
100 V, 1 A
100 V, 1 A
100 V, from 0.5 A to 1 A
1.5 mA, 100 V
1 A, 100 V
1 A, 100 V