Agilent Technologies N6751A User Manual page 120

Modular power system
Table of Contents

Advertisement

Appendix A Specifications
N6731B/
N6741B
DC Output Ratings:
Voltage
Current
10 A / 20 A
NOTE 1
Power
50 W / 100 W
Output Ripple and Noise (PARD):
(from 20 Hz – 20 MHz)
CV peak-to- peak
10 mV / 11mV
CV rms
Load Effect (Regulation):
(with output change from no load to full load, up to a maximum load-lead drop of 1V/lead)
Voltage
Current
Source Effect (Regulation):
Voltage
Current
Programming Accuracy:
(@ 23 °C ±5 °C after 30 minute warm-up. Applies from minimum to maximum programming range)
Voltage
0.1% + 19 mV
Current
0.15% + 20 mA
Measurement Accuracy:
(at 23 °C ±5 °C)
Voltage
0.1% + 20 mV
Current
0.15% + 20 mA
Load Transient Recovery Time:
(time to recover to within the settling band following a load change
from 50% to 100% and from 100% to 50% of full load.)
Voltage settling band
± 0.08V / 0.1 V
Time
< 200 µs
1
Output current is derated 1% per °C above 40°C.
120
Agilent Models N6731B - N6736B and N6741B - N6746B
N6732B/
N6742B
5 V
8 V
6.25 A / 12.5 A
50 W / 100 W
12 mV
2 mV
2 mV
5 mV
6 mV
2 mA
2 mA
1 mV
2 mV
1 mA
1 mA
0.1% + 19 mV
0.15% + 20 mA
0.1% + 20 mV
0.15% + 10 mA
± 0.08V / 0.1 V
< 200 µs
N6733B/
N6734B/
N6743B
N6744B
20 V
35 V
2.5 A / 5 A
1.5 A / 3 A
50 W / 100 W
52.5W / 105W
14 mV
15 mV
3 mV
5 mV
9 mV
11 mV
2 mA
2 mA
2 mV
4 mV
1 mA
1 mA
0.1% + 20 mV
0.1% + 35 mV
0.15% + 20 mA
0.15% + 20 mA
0.1% + 20 mV
0.1% + 35 mV
0.15% + 5 mA
0.15% + 4 mA
± 0.2 V / 0.3 V
± 0.2 V / 0.3 V
< 200 µs
< 200 µs
N6735B/
N6736B/
N6745B
N6746B
60 V
100 V
0.8 A / 1.6 A
0.5 A / 1 A
50 W / 100 W
50 W / 100 W
25 mV
30 mV
9 mV
18 mV
13 mV / 16 mV
20 mV / 30 mV
2 mA
2 mA
6 mV
10 mV
1 mA
1 mA
0.1% + 60 mV
0.1% +100 mV
0.15% + 20 mA
0.15% + 10mA
0.1% + 60 mV
0.1% +100 mV
0.15% + 4 mA
0.15% + 2 mA
± 0.4 V / 0.5 V
± 0.5 V / 1.0 V
< 200 µs
< 200 µs
Series N6700 User's Guide