Specifications; Installation; Pre-Installation/Installation; Required Tools & Accessories - Turner Designs TD-107 5.0 Operation Manual

Oil content monitor
Table of Contents

Advertisement

3.0 SPECIFICATIONS

NAG Marine Part Number: NAG2541
Physical Specifications
17.375 in. W x 7.1875 in. D
Dimensions
x 15.0625 in. H [442.1 mm
(overall):
W x 182.4 mm D x 383
mm H]
Dimensions
9 in W x 12.25 in H
(enclosure
228.6 mm W x 311.2 mm H]
only):
Weight:
12 lbs. [5.4 kg]
Enclosure
FRP & Stainless Steel
Standards:
NEM 4
Inlet Valve Pipe
Size:
0.25 in. [6.35 mm] FNPT
Outlet Pipe Size: 0.25 in. [6.35 mm]
compression tube fitting
Electrical Specifications
20-240 VAC/VDC, 50/60
Power:
Hz, 15 watts
Dry contact, 240 VAC or 28
Alarm Relays (5
VDC, 6 A maximum, Normally
total):
Open/Normally Closed
connections
Continuity or 3–28 VDC, 10
System Status
mA minimum, 100 mA
Inputs (2 total):
maximum
Signal Output (1
Internally powered, 4–20 mA
total):
(isolated)
Data Retrieval:
One internal USB port.
Operating Limits
100 psig [689.5 kPa(g)]
Sample Pressure:
maximum
Sample
140 °F [60 °C] maximum
Temperature:
Ambient
40–120 °F [4–49 °C]
Temperature:
Relative
Humidity
0–100 %
Range:
TD-107
Oil Types
Measured:
Oil
Measurement
Range:
Accuracy oil +
solids:

4.0 INSTALLATION

4.1 Pre-Installation/Installation

Do not break the provided
tamper evident DNV certification
seal for any reason. Breaking
such seals will invalidate the DNV
certification of the monitor.
The Pre-Installation/Installation Checklist
found in Appendix A provides important
guidelines and information to aid in
preparing for installation.
4.2 Required Tools & Accessories
Standard plumbing and electrical tools
are required for installation.
NAG Marine recommends that stainless
steel flexible tubing be used on the inlet
and outlet connections. This will
facilitate future maintenance and
prevent damage to the monitor from to
transmitted vibration and/or pipe strain.
DNV Certified OWS system including
diverter valve.
4.3 Location & Sampling Point
The sampling point should be located such
that the sample is homogeneous and free of
all air. It should also approximate the bulk
oil concentrations of the treated bilge water.
It is extremely important to eliminate air
entrapment in the sample line. The best way
to accomplish this is to sample from the
center of the pipe or from the side of the
pipe.
7
5.0
Oil Content Monitor
Operation Manual
As per DNV TAP No.
771.61 requirements.
Scaled 0–30 ppm, as per
DNV TAP No. 771.61
requirements.
As per DNV TAP No.
771.61 requirements.

Hide quick links:

Advertisement

Table of Contents
loading