Summary of Contents for Turner Designs C-FLUOR

  • Page 1 User’s Manual April 8, 2019 P/N 998-2123 Revision 1.0 TURNER DESIGNS 1995 N. 1 Street San Jose, CA 95112 Phone: (408) 749-0994 FAX: (408) 749-0998...
  • Page 2: Table Of Contents

    Warranty Service Out of Warranty Service Appendices Specifications Recommended Lab Measurement Practices Wiring Guide Pigtail Cable and Connector Information Linear Range, Quenching and Temperature Considerations Functional Test for C-FLUOR with No Connector Using the C-FLUOR Submersible Probe with the DataBank...
  • Page 3 WASTE ELECTRICAL AND ELECTRONIC EQUIPMENT (WEEE) DIRECTIVE Turner Designs is in the business of designing and selling products that benefit the well-being of our environment. Accordingly, we are concerned with preserving the surroundings wherever our instruments are used and happy to work with customers by complying with the WEEE Directive to reduce the environmental impact resulting from the use of our products.
  • Page 4: Introduction

    C-FLUOR probe. The certificate contains calibration values that are used to convert voltage to concentration estimates. C-FLUOR is manufactured with a Titanium housing and connector, ideal for long-term deployments and able to withstand most corrosive environments. It is available without a connector for integration flexibility;...
  • Page 5: Inspection And Setup

    C-FLUOR Submersible Probe 2. Inspection and Setup 2.1 Instrument Checklist The C-FLUOR Submersible Probe shipment package consists of:  C-FLUOR Submersible Probe: Configured and factory calibrated for the specified fluorophore or material as noted by the Identification Letter stamped on the connector: ...
  • Page 6 Brighteners, and Crude Oil P/N: 2100-904 - see Accessory Instructions on USB Flash Drive Note: The UV SSS can be used for both C-FLUOR and Cyclops for the applications specified above, as well as Cyclops Refined Fuels and Tryptophan sensors.
  • Page 7: Functional Test For C-Fluor With Connector

    Pigtail cable holes Analog Ground “-” (Green) With the C-FLUOR connected as shown in Figure 1 answer questions 1-3 by making the following functional tests: 1. Is the LED on? Hold a piece of white paper about ½ an inch in front of the optical head to ensure the LED is ON.
  • Page 8: Measurements With The C-Fluor

    Section 4.1 for information regarding maintenance. C-FLUOR Probes output a voltage signal that is proportional to the fluorophore of interest. Every C-FLUOR is factory calibrated and delivered with a certificate that contains factory calibration values for that specific probe; the certificate also contains an equation that can be used with the calibration values to convert C- FLUOR’s voltage signal to concentration estimates for the specified fluorophore...
  • Page 9: Performing C-Fluor Measurements

    C-FLUOR Probes output a 0-5 Volt analog signal proportional to the fluorescence detected from the fluorophore of interest. Use the wiring diagram from Section 2.4 to connect the C-FLUOR to a power supply and a multi-meter that will read voltages from the probe. Submerge the optical head, or entire probe, into the water sample;...
  • Page 10: Maintenance And Warranty

    4. Maintenance and Warranty 4.1 Maintenance 4.1.1 Rinsing The C-FLUOR should be rinsed or soaked in fresh water following each deployment until it is completely clean again. 4.1.2 Care for the bulkhead connector A light coat of Silicone spray should be used on the rubber of the male pins of the bulkhead to aid in sealing.
  • Page 11: Warranty Service

    4.3 Warranty Service To obtain service during the warranty period, the owner shall take the following steps: 1. Write, email or call Turner Designs Technical Support and describe as precisely as possible the nature of the problem. Phone: 1 (877) 316-8049 Email: support@turnerdesigns.com...
  • Page 12 C-FLUOR Submersible Probe 998-2123 Rev. 1.0 Page 12...
  • Page 13: A Specifications

    C-FLUOR Submersible Probe Appendix A: Specifications Parameter Specification Full Range Linearity 0.99 R CHL, RWT, PC, PE, F, CDOM/fDOM, Oil, OB = 12mA Power Draw @ 12V Turbidity = 10mA 3 – 15 VDC Input Voltage 0 – 5 VDC Analog Signal Output Ambient: 0 to 50 C...
  • Page 14: B Recommended Lab Measurement Practices

    C-FLUOR Submersible Probe Appendix B: Recommended Lab Measurement Practices Recommended Lab Practices for Measurements The following steps will improve the accuracy and repeatability of your measurements, especially at low concentration levels: 1. Use a non-fluorescent container for your water samples. Note: Plastic may fluoresce and interfere with the sample’s fluorescence.
  • Page 15: C Wiring Guide

    C-FLUOR Submersible Probe Appendix C: Wiring Guide C-FLUOR Function Connection Pins/Wires Number Supply Voltage PSU – Positive Connection 3 – 15 VDC PSU – Ground Connection Black Supply Ground, 0VDC Signal Out to data logger “+” Multimeter Positive Orange 0 – 5VDC Connection Analog Ground “-”,...
  • Page 16: D Pigtail Cable And Connector Information

    Wire Color A maximum cable length up to 100 meters can be connected to the Cyclops if the A maximum cable length up to 100 meters can be connected to the C-FLUOR if the following conditions are met: following conditions are met:...
  • Page 17: E Linear Range, Quenching And Temperature Considerations

    For further information on how temperature, light, water quality and the physiological state of the algal cells can all affect measurements. Please refer to the Frequently on Turner Designs’ website for more Asked Questions in Technical Support information about temperature compensation.
  • Page 18: F Functional Test For C-Fluor With No Connector

    Note: Yellow and Blue wires are not used. Analog Ground “-” (Green) With the C-FLUOR connected as shown in Figure 1 answer questions 1-3 by making the following functional tests: 4. Is the LED on? Hold a piece of white paper about ½ an inch in front of the optical head to ensure the LED is ON.
  • Page 19: G Using The C-Fluor Submersible Probe With The Databank

    Profiling – purchase extended cables to allow for vertical profiling Monitoring – deploy the C-FLUOR to a fixed location or depth and monitor the signal over time; set up logging to capture a signal within a specific time frame; download data while the probe is deployed and continuously measuring;...