4.6.2
Disconnecting Wiring Connections
To disconnect a wire from the pre-wired
cabling:
1. Turn off main power to the instrument at
the breaker.
2. Verify all process inputs and switched
outputs are off.
3. Access the junction box to disconnect
the wire.
Signal wires and power wiring
should not be run in the same
conduit. Failure to separate or shield
these wires can result in electrical
interference.

5.0. START-UP / SHUT-DOWN

5.1 Start-Up Materials

5.0
The TD-107
OCM (P/N NAG2541)
includes the following components:
TD107™ OCM Instrument Assembly
2-Year Spares Kit: P/N NAG2520
Operation Manual: P/N NAG2544
Components and materials supplied by the
user:
DNV Certified Oily Water Separator with
Diverter Valve
Deionized (DI) or distilled water
Tubing and fittings for mechanical
connections are supplied by the installer.
If necessary, please contact your NAG
Marine representative or contact NAG
Marine at +1 757 852 3998 for additional
information.

5.2 Start-Up Procedure

Diagrams of the screens are
located in Appendix E.
Perform the following steps to start the TD-
5.0
107
OCM:
1. Bring sample flow to and from the TD-
1075.0 OCM; check for leaks in
plumbing.
2. Apply power. When power is first
turned on, an identification screen
will appear for a few seconds,
showing the software version and
date. After 10 seconds, the HOME
screen will appear.
screen will display the oil content.
Sample HOME Screen:
3.5 PPM
TD-107
If the screen appears blank or dark,
try adjusting the screen contrast
using the UP <
arrows.
The user must return to the HOME
screen (by pressing either <HOME>
or <ESC>) after verifying each
value. Jumping between screens without
returning to the HOME screen is not
permitted.
3. During start-up, verify the following
settings:
a.
High Signal Alarm Concentration
Press <1> to verify the factory
programmed value of 5.0 ppm.
b.
Early Warning
Press <2> to verify the programmed
value. This value may be user set.
c.
Zero Signal Alarm
Press <3> to verify the factory
programmed value of 0.0 ppm.
d.
4 mA Output
Press <4> to verify the factory
programmed value of 0.0 ppm.
e.
20 mA Output
Press <6> to verify the factory
programmed value of 30.0 ppm. This
value may be user set to adjust for
existing equipment.
The User ID will intentionally not
allow access to features of the
monitor which would compromise its
application performance or data storage.
10
The HOME
5.0
> and DOWN <
>