FluoroSense™
User's Manual
December 21, 2018
P/N: 998-2852
Revision C
TURNER DESIGNS
st
1995 N. 1
Street.
San Jose, CA 95112
Phone: (408) 749-0994
FAX: (408) 749-0998

Advertisement

Table of Contents
loading

Summary of Contents for Turner Designs FluoroSense

  • Page 1 FluoroSense™ User’s Manual December 21, 2018 P/N: 998-2852 Revision C TURNER DESIGNS 1995 N. 1 Street. San Jose, CA 95112 Phone: (408) 749-0994 FAX: (408) 749-0998...
  • Page 2: Table Of Contents

    Introduction Description Inspection and Setup General Information, Precautions & Cleaning Instrument Operation and Calibration Instrument Power Up Sample Analysis Calibration Procedure (only needed if FluoroSense fails calibration check) Applications In Vivo Chlorophyll Warranty Terms Warranty Service Out of Warranty Service...
  • Page 3 WASTE ELECTRICAL AND ELECTRONIC EQUIPMENT (WEEE) DIRECTIVE Turner Designs is in the business of designing and selling products that benefit the well-being of our environment. Accordingly, we are concerned with preserving the surroundings wherever our instruments are used and happy to work with customers by complying with the WEEE Directive to reduce the environmental impact resulting from the use of our products.
  • Page 4: Introduction

    Introduction 1.1 Description The FluoroSense is a lightweight, handheld fluorometer ideal for field use. Its small size makes it perfect for the user who needs quick measurements away from the laboratory. 1.2 Inspection and Setup 1.2.1 Inspection Upon receiving your instrument, please inspect everything carefully and make sure all accessories are present.
  • Page 5: Instrument Operation And Calibration

    The display will flash “td” several times and then the unit will automatically turn off. At this point the FluoroSense is calibrated. You can confirm the calibration by turning the unit on and pressing the read button while the unit is in the standard solution. Readings should be 100 ±...
  • Page 6: Applications

    µg/L of PC. The conversion factor was determined using a known concentration of extracted PC in buffer solution, diluted in water. The actual concentration estimated by the FluoroSense is semi-quantitative at best as extracted PC has a different fluorescence yield from PC contained within live algal cells.
  • Page 7: Terms

    (RMA). Package the unit, write the RMA number on the outside of the shipping carton, and ship the instrument, prepaid, to Turner Designs. If the failure is covered under the warranty terms, the instrument will be repaired and returned free of charge, for all customers in the contiguous continental United States.
  • Page 8: Out Of Warranty Service

    Shipment to Turner Designs should be prepaid. Your bill will include return shipment freight charges. Address for Shipment: Turner Designs 1995 N. 1 Street San Jose, CA 95112 FluoroSense™ User’s Manual...
  • Page 9 FluoroSense™ User’s Manual...
  • Page 10: Appendixes A Specifications

    4.3 oz; 120 g Temperature 41-104°F (5-40°C) Case IP 67 standard; dustproof/waterproof Power 2 AA batteries, standard or rechargeable Auto Power Off After 1 minute of inactivity Temperature coefficient for in vivo chlorophyll 1.4%/ °C or .78%/ °F FluoroSense™ User’s Manual...
  • Page 11: B Linear Range, Quenching And Temperature Considerations

    Appendix B: Linear Range, Quenching and Temperature Considerations Linear Range and Quenching The linear range is the concentration range in which the readout of the FluoroSense is directly proportional to the concentration of the fluorophore in the sample. The linear range begins with the lowest detectable concentration and spans to an upper limit concentration that is dependent on the properties of the fluorescent material and the cuvette optical path length.