7. Inject another 50ml of clean fresh water into the sample cell and observe the display screen.
8. On the display screen, you will see (a) Alarm, (b) PPM value, (c) Cell Temperature, (d) Cell
Condition Percentage (between 0% ‐ 100%).
9. Allow PPM value to stabilize (typically between 0‐5ppm in fresh clean water).
10. Allow the Cell Condition Percentage to stabilize (typically between 70‐100% in fresh clean water).
11. If the PPM value has stabilized between 0‐5ppm AND the Cell Condition
percentage has stabilized between 70‐100%, the unit is ready for
operation.
12. Place the sample cell valve handle into Operating position and begin OWS operations.
13. If the PPM value OR Cell Condition percentage is still out of starting
parameters, repeat the acidic cleaning process.