Turner Designs TD-107 5.0 Operation Manual page 48

Oil content monitor
Table of Contents

Advertisement

5.0
TD-107
Oil Content Monitor
Operation Manual
To Change:
Early Warning
2
Alarm Value
User ID
5
F.1. System Value Definitions
F.1.1. High Std Conc Value
This value relates the fluorescence
measurement value to a high standard
concentration reference oil content
fluorescence measurement.
F.1.2. Low Std Conc Value
This value relates the fluorescence
measurement value to a low standard
concentration reference oil content
fluorescence measurement.
F.1.3. High (5 ppm) Signal Alarm Value
This is the value at which the High (5
ppm) Signal Alarm is triggered.
F.1.4. Early Warning Value
This is the value at which the Early
Warning is triggered.
F.1.5. Zero Signal Alarm Value
This is the value at which the Zero Signal
Alarm is triggered.
Press the keys in this order:
Enter desired value
ENT
(1.0–4.9)
Master
ENT
ENT
ID
ENT
ESC
Enter
desired
ENT
ESC
value.
F.1.6. 4 mA Output Value
The lower limit of the analog output sent
to the Datalogger, which is 0 ppm.
F.1.7. User ID / Master ID
The User ID allows the user to access
certain functions in the software not
accessible to those who do not possess
the User ID. The Master ID allows access
to the same functions as the User ID and
includes others such as the ability to
change the User ID.
F.1.8. 20 mA Output Value
The upper limit of the analog output sent
to the Datalogger, which is 30 ppm.
The date and time values are
accessed under the <> menu.
The Datalogger date and time are
set to Greenwich Mean Time (GMT)
regardless of the date and time setting in
5.0
the TD-107
cannot be changed.
48
OCM. The time and date