5.0
TD-107
Oil Content Monitor
Operation Manual
Be sure to move the inlet valve to
the flow position after completing the
cleaning.
9.2.3
Calibration Check
The calibration should be checked using the
provided Check Solution. If the Check
Solution does not read within the allowed
tolerance, clean the sample cell. If the
calibration discrepancy persists, contact the
factory-authorized dealer. The TD-107
OCM must be re-calibrated on a periodic
basis as indicated on the calibration
certificate. Per DNV TAP No. 771.61
regulations, this calibration must be
conducted by the factory-authorized dealer.
Contact the factory-authorized dealer within
the two months prior to the expiration of the
calibration certificate to arrange re-
calibration.

9.3 Replacement Parts

5.0
TD-107
OCM replacement parts can be
obtained from your NAG Marine
representative or by calling NAG Marine at
+1 757 852 3998.
Common Replacement Parts:
NAG2544 Operation Manual (Serial
numbers 2000001F and higher)
NAG2504 60 mL Syringe (Qty. 3)
NAG2505 Desiccant Plugs (Qty. 3)
NAG2506 Flow-cell Brushes (Qty. 3)
NAG2508 Cap, Carboy w/Luer-Lok®
connector
NAG2509 Cap, Calibration, w/Luer-Lok®
connector
NAG2510 Carboy, 1 Gallon
NAG2520 2-Year Spares Kit
(includes desiccant plug, flow
cell brush, cleaning solution,
check solution, and syringes)
NAG2522 Cleaning Solution
NAG2528 Cleaning Spares Kit
NAG2529 Calibration Check Spares
NAG2530 Clean Out Plug Replacement
Fitting
NAG2532 Check Solution
5.0
20