Turner Designs FluoroSense Quick Start Manual

Turner Designs FluoroSense Quick Start Manual

Handheld fluorometer
Hide thumbs Also See for FluoroSense:

Advertisement

Note: Unit ships calibrated. Only
re-calibrate if unit does not read
100
appropriate for your FluoroSense.
• To calibrate, press and hold Read
button while unit is off.
• While holding Read button, press and
hold Power button until 100 appears
on screen. Release both buttons.
• Place the sensor in the FluoroSense
Standard Solution appropriate for
your optics. Shake gently to make
sure bubbles are purged from the cap
and press Read button.
• The display will flash "td" several
times and then the unit will
automatically turn off. At this point
the FluoroSense is calibrated.
998-2851

Battery Installation

Calibration

3% for the Standard Solution
READ
www.turnerdesigns.com
Quick Start Guide
100
B
Page 1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the FluoroSense and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Turner Designs FluoroSense

  • Page 1 • While holding Read button, press and hold Power button until 100 appears on screen. Release both buttons. • Place the sensor in the FluoroSense Standard Solution appropriate for your optics. Shake gently to make sure bubbles are purged from the cap and press Read button.
  • Page 2 • If there is any noticeable fouling, use a soft bristle brush or non-abrasive cloth and soapy water to clean the optical window. Rinse thoroughly. Note: The FluoroSense should NOT come in contact with any organic solvents or strong acids and bases. www.turnerdesigns.com...

Table of Contents