3. Remove the cartridge or container, as instructed by the wizard, 
confirm this in the wizard screen, and click 
Figure 6-87: Confirmation of cartridge removal
4. In the next wizard screen, observe the numbers and wait until the level 
is relatively stable—two units above or below the average level shown.
Figure 6-88: Level stabilization
5. Select the check box, and click 
If you need to calibrate more than one load cell, the next calibration 
screen appears. Repeat steps 4 and 5 until all of the load cells are 
calibrated.
6. Click 
 in the final wizard screen.
Done
Figure 6-89: Final wizard screen