Setting Of Alarm Values - Fuji Electric ZPG Instruction Manual

Ndir type infrared gas analyzer
Hide thumbs Also See for ZPG:
Table of Contents

Advertisement

6.3 Alarm setting

6.3.1 Setting of alarm values

ENT
key.
CAUTION
Set the values so that H-limit value > L-limit
value and that (H-limit value - L-limit value)
> hysteresis.
When "0" is set, the alarm operation is not
performed.
To close the "Alarm Setting"
To close the "Alarm Setting" or to cancel this
mode midway, press the
A previous screen will return.
Setting range
0% to 100% FS (Settable in each range).
INZ-TN2ZPG-E
key. Press
ENT
key.
ENT
key.
ESC
key.
45
Alarm Setting
Select Alarm No. or
Hysteresis setting
Alarm-1
Alarm-2
Alarm-3
Alarm-4
Alarm-5
Alarm-6
Hysteresis
0 0
Alarm Setting
Select an item
with UP/DOWN and ENT
Alarm-1
Back with ESC
Channel
H-Limit Range 1
Range 2
L-Limit Range 1
Range 2
Kind of Alarm
ON / OFF
Cursor for setting value
Alarm Setting
Set value
Alarm-1
Channel
H-Limit Range 1
Range 2
L-Limit Range 1
Range 2
Kind of Alarm
ON / OFF
End of Alarm Setting
%FS
(
)
ENT
Ch1
10.00
ppm
100.0
ppm
000.0
ppm
0000
ppm
High
OFF
ENT
(
)
Ch1
10.00
ppm
100.0
ppm
000.0
ppm
0000
ppm
High
OFF
ENT
(
)