Functional Checkout procedures

Frequency

Identify enabled patient parameters and software options

8-20
Functional and electrical safety checks: Functional Checkout procedures
Qualified personnel must perform the checkout procedures:
Every 12 months after receipt of the device (Preventative
Maintenance). Refer to
information.
Each time the main enclosure is disassembled or a circuit board is
removed, tested, or replaced (Corrective Maintenance). Refer to
"Recommended checkout"
The check out procedures support all enabled patient parameter and
software options. To view the options enabled on the patient monitor, go
to the Main Menu and select MORE MENUS > MONITOR SETUP >
SOFTWARE CONFIGURATION.
The following is a list of all available patient parameter and software
options.
0BP (invasive)
1BP or 2BP (invasive)
3BP or 4BP (invasive)
12SL
ACI-TIPI
HI-RES TRENDS
ETCO
2
NETWORK
AVOA Plus
CARDIO-PULMONARY
CARDIAC
Dash 3000/4000/5000
"Maintenance schedule"
on page 7-52 for more information.
(CO
) (End-tidal CO
)
2
2
on page 5-2 for more
2000966-542D