Noninvasive Blood Pressure Tests - GE Dash 3000 Service Manual

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Noninvasive blood pressure tests

Equipment
2000966-542D
Functional and electrical safety checks: Functional Checkout procedures
12. Press and release the 8h.
13. Switch and verify the SRC-MAX default indicators are lit as follows:
HEART RATE = 200 bpm
LIGHT = HIGH
%SpO2 = 90
MODULATION = HIGH
Allow the Dash a few seconds to obtain a steady reading.
14. Verify the following SpO2 readings on the patient monitor for
saturation and pulse rate:
Saturation (%): 90 ±2
Rate (bpm): 200 ±3% (194 to 206)
Digital Manometer with a range of at least 0 to 1000 mmHg Accuracy
0.5% FS.
NBP cuff coupling (400787-005 or equivalent)
NBP hose coupling (400787-006 or equivalent)
NBP tee (4745-101 or equivalent)
NBP tubing 2 feet (401582-001 or equivalent)
Manometer: mercury, digital (Sensym PDM200M – no longer
available for ordering, Meriam Instrument Smart Manometer Model
350 DM2000, or equivalent)
NBP tube (2017008-001 or equivalent)
NBP cuff (2203 or equivalent)
Pipe: PVC
Dash 3000/4000/5000
8-35

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dash 4000Dash 5000001c 051d 003a

Table of Contents