GE Dash 3000 Service Manual page 25

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Optional alarm light indicator
2000966-542D
Equipment overview: Components
Name
Defib Sync port
E
Aux port
F
Ethernet port
G
peripheral expansion
H
port
AC power connector
I
equipotential terminal
J
product code label
K
An optional alarm light indicator may be built into the handle of the
Dash 3000 patient monitor or into the display bezel of the Dash 4000/
5000 patient monitor. When activated, the LED indicator flashes red for
Dash 3000/4000/5000
Description
Provides ECG analog output signals to user-supplied
equipment. A 5-volt, 2-millisecond artificial pacer spike
is added to the analog output when PACE is on and
detection occurs.
Provides serial connection to compatible auxiliary
devices, including:
Remote control
TRAM-RAC housing
Remote display
®
Nellcor
395 pulse oximeter
®
Aspect
BISx module
Used to connect a monitor to the Unity Network for
patient monitoring or for software installation.
Used for connecting to a Dash™ Port docking station or
other compatible auxiliary devices.
Provides connection for AC power cable. The internal
power supply converts the AC current to DC power.
The monitor is preset at the factory for a specific AC
voltage. Refer to the label on the back of the unit for the
voltage and current requirements. Before applying
power, check that the power requirements match your
power supply.
For measurements in or near the heart we recommend
connecting the monitor to the potential equalization
system. Use the green and yellow potential equalization
cable and connect it to this pin.
Identifies the product code for this monitor for
identification and service needs.
2-5

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dash 4000Dash 5000001c 051d 003a

Table of Contents