Cautions - GE Dash 3000 Service Manual

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Cautions

5-4
Preventive maintenance: Cleaning and disinfecting the patient monitor
Any sodium hypochlorite wipe product that meets the above
guidelines can be used.
NOTE
Wring excess disinfectant from wipe before using.
NOTE
Any contact of the disinfectant solutions with metal parts may
cause corrosion.
3. When disinfecting the equipment, allow disinfectant solution to
remain on device for a minimum of one minute per hospital
guidelines.
4. Wipe off cleaning solutions with a clean, lightly moistened cloth.
5. Dry thoroughly with a dry lint-free cloth and let air dry for at least
30 minutes.
NOTE
Drying times may vary based on the environmental conditions.
6. Take care not to let fluid "pool" around connection pins. If this should
happen, blot dry with a soft lint-free cloth.
Use the following precautions to prevent damaging the equipment:
Never immerse the device in any liquid.
Do not pour or spray any liquid directly on the device or permit fluid
to seep into connections or openings.
Never use conductive solutions, solutions that contain chlorides, wax,
or wax compounds to clean the device.
Never use solutions or products that contain the following:
Any type of Ammonium Chloride such as, but not limited to
Dimethyl Benzyl Ammonium Chloride and Quaternary
Ammonium Chloride solutions.
Abrasive cleaners or solvents of any kind.
Acetone
Ketone
Betadine
Alcohol-based cleaning agents
Sodium salts
Never autoclave or steam clean the device.
Always use household glass cleaner to clean the touchscreen or
standard displays.
Dash 3000/4000/5000
2000966-542D