Tool Requirements; Maintenance And Checkout Procedures - GE Giraffe Blue Spot PT Lite Operation, Maintenance And Service Manual

Table of Contents

Advertisement

Chapter 8: Service Maintenance and Checkout

8.3 Tool Requirements

The following table lists the service tools required to perform the planned maintenance and checkout
procedures:
Checkout Procedure Name
"8.4.1 Visual Inspection"
"8.4.2 Functional Checks"
"8.4.3 Light Intensity Check"
"8.4.4 Electrical Safety Tests"

8.4 Maintenance and Checkout Procedures

NOTE: For instructions on cleaning the unit cover, refer to Chapter 3.
8.4.1 Visual Inspection
1. Disconnect the power cord from the power outlet.
2. Examine the power cord for any signs of damage. Replace the cord if damage is evident and perform
electrical safety test as instructed in section 8.4.4.
3. Check that the power cord retainer is in place.
4. Examine the system overall for any damaged (cracked or broken) or missing parts, such as the plastic
unit cover, standby/on switch, standby/on indicator, over/under temperature indicator, hour meter
display, light pipe, light shade, air filter guard, or potential equalization stud. If any part is damaged or
missing, replace it (Refer to Chapter 11).
5. Examine the air filter for damage or dust. Replace if necessary.
6. Examine the system for any missing or damaged labels. For a list of labels, and for the proper location
of each label, refer to section 12.2.
7. For units on portable roll stand, examine the roll stand for any damaged or missing parts. Check the
wheel locks of the roll stand and confirm that they are functional.
52
Table 8-2: Tool Requirements
Service Tools Needed
None
None
BiliBlanket Light Meter II (6600-0198-900),
Tape measure
Safety analyzer
Quantity Needed
NA
NA
1
1