2000966-542D
Field replaceable units: Remove or replace handle assembly
Carefully disconnect the antennas using a pen tip, small
screwdriver, or needle nose pliers.
Place the wireless card in a safe place.
Remove the handle from the main unit, carefully removing
the antenna cables from the housing holes.
d. To replace the handle, route the antenna cables through the
housing holes as shown. Apply adhesive tape to temporarily hold
the handle into position.
e. Gather the antenna cables together in the middle of the unit,
then do the following:
Make sure that the cables are lying flat and are not twisted
around each other.
Provide enough slack in the cables to allow the handle to be
removed if future repairs are required.
Remove the protective backing from the short antenna label
and position the label over the untwisted antenna cables as
shown.
Dash 3000/4000/5000
942A
954B