12SL and ACI-TIPI ECG test (option)
8-24
Functional and electrical safety checks: Functional Checkout procedures
Perform this test if your patient monitor uses the 12SL ACI-TIPI ECG
option.
1. Set up the patient simulator as follows:
Heart rate – 80 bpm
Heart rate amplitude – 1.0 mV
12SL ECG patient cable (5-leadwires with V leadwires) properly
attached.
2. Select ECG from the patient monitor menu. Then, select 12 LEAD
ECG ANALYSIS.
3. Verify that the patient monitor is displaying 12 noise-free leads.
4. Select 12LD ECG NOW. Wait for the patient monitor to acquire and
analyze the data.
5. Select Transmit-Print.
6. Verify the 12SL ECG prints at the print location assigned in the
patient monitor's Print Setup >12SL Print Location.
If there is no print location is assigned, an error message appears
on the bottom of the patient monitor's display.
7. Verify the ECG is transmitted to the MUSE Cardiovascular
Information System. Verify the ECG prints out correctly as defined
by the MUSE system.
If no MUSE system is connected, an error message appears on
the bottom of the patient monitor's display.
8. Delete this test 12SL ECG from the MUSE system's edit list.
Dash 3000/4000/5000
2000966-542D