Test pattern (option based on Dash having a local printer)
Graph speed

Display test

2000966-542D
Functional and electrical safety checks: Functional Checkout procedures
Using the Trim Knob control, access the Service Mode menu starting
from the Main Menu.
1. Select MORE MENUS > MONITOR SETUP > SERVICE MODE.
2. Enter password using the Trim Knob control to select the day and
month from patient monitor screen with leading zeros. (e.g. July 4 =
0407).
3. Select PRINT (or GRAPH) TEST PATTERN > START.
4. Verify the following:
Fonts
Shading
Triangle Pattern
No missing dots
5. Select PRINT (or GRAPH) TEST PATTERN > STOP.
Using the Trim Knob control, access the Graph Setup menu starting
from the Main Menu.
1. Select MORE MENUS > MONITOR SETUP > PRINT (or
GRAPH) SETUP.
2. Select SPEED:25 (default).
3. Verify that all eight speeds work.
1. Hold the NBP Auto and the ZERO ALL keys and press the Trim
Knob control at the same time.
2. Release the Trim Knob control immediately.
3. Continue holding the NBP Auto and the ZERO ALL keys.
4. Select "Video Test Screens."
5. Test all screens:
White Screen
Red Screen
Blue Screen
Green Screen
Vertical Bars
Dash 3000/4000/5000