Replace Power Supply Assembly - GE Dash 3000 Service Manual

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Replace power supply assembly

7-48
Field replaceable units: Replace main unit parts
a. Connect the antenna cables to the wireless card (black cable to
the left and grey cable to the right).
b. With the antenna connection facing down, install the card into
the card bracket.
13. Reassemble the patient monitor in reverse order.
14. Be sure to complete the following procedures in
checkout"
on page 7-52:
electrical safety tests
checkout procedures
calibration
15. Also configure the patient monitor and enable the software options
previously installed. Refer to
1. Remove the CPU/battery housing assembly according to steps
starting on page 7-44.
2. While holding the power supply assembly with one hand, remove the
four screws from the back of the unit.
Dash 3000/4000/5000
Black antenna cable (left)
Grey antenna cable
(right)
Antenna connectors
"Recommended
"Configuration"
on page 4-1.
959C
575A
2000966-542D