Prompts alert the user but may not stop the infusion and are indicated by an audible sound, a flashing amber warning indicator and a message on the
display.
1. Check the display for a prompt message. Press
2. Rectify the cause of the prompt or proceed with caution.
Prompts:
Display
ATTENTION
SET VTBI
SET NOT FITTED
LOCKED
Advisories:
Display
DOSE WOULD EXCEED
DOSE UNDER
DOSE NOT PERMITTED
RATE NOT PERMITTED
CONCENTRATION NOT
PERMITTED
WEIGHT ABOVE LIMIT
WEIGHT BELOW LIMIT
Prompts and Advisories
to silence the sound for 2 minutes, CANCEL to clear the message.
c
Infusion
Cause
Status
Pump on hold
Pump left on hold for 2 minutes
without starting the infusion.
Pump on hold
No VTBI / flow sensor.
Pump on hold
No infusion set fitted.
Infusion continues
Rate change attempted whilst
locked.
Infusion
Cause
Status
Pump on hold (If
Infusion rate set exceeds a
titration is disabled)
Guardrails® soft limit.
Infusion continues (If
titrating)
Pump on hold (If
Infusion rate/dose rate set is
titration is disabled)
under a Guardrails® soft limit.
Infusion continues (If
titrating)
Pump on hold (If
Dose rate entered is greater than
titration is disabled)
the dose rate hard maximum set.
Infusion continues (If
titrating)
Pump on hold (If
Infusion rate set exceeds a
titration is disabled)
Guardrails® hard limit.
Infusion continues (If
titrating)
Pump on hold
Concentration set exceeds
hard max limit, or is under hard
minimum limit.
Pump on hold
Patient weight set exceeds a
Guardrails® soft limit.
Pump on hold
Patient weight set is under a
Guardrails® soft limit.
1000DF00498 Issue 3
Action
• Review pump setup.
• Start infusion or turn off pump.
• Set VTBI or fit flow sensor.
• Fit infusion set.
• Unlock rate to adjust infusion settings.
Action
• Check infusion setting.
• To confirm OVERRIDE LIMIT? press YES.
• To deny OVERRIDE LIMIT? press NO.
• Check infusion setting.
• To confirm OVERRIDE LIMIT? press YES.
• To deny OVERRIDE LIMIT? press NO.
• Check infusion setting and adjust to
appropriate required rate.
• Check infusion setting and adjust to
appropriate required rate.
• Check concentration and adjust to a more
appropriate amount.
• Check weight setting.
• To confirm CONFIRM? press YES.
• To deny CONFIRM? press NO.
• Check weight setting.
• To confirm CONFIRM? press YES.
• To deny CONFIRM? press NO.
27/42