Clearing the drug name is only available if drug name only has been selected:
h
to put the pump ON HOLD.
1. Press
d
button to access the options menu.
2. Press the
3. Select DRUG NAME ONLY using the
4. Select CLEAR DRUG NAME (displayed if a name only is selected) using the
If a secondary infusion has already been setup (see 'Secondary (Piggyback) Infusions' section), then access to the primary infusion setup is as
follows:
h
to put the pump ON HOLD.
1. Press
d
button to access the options menu.
2. Press the
3. Select PRIMARY SETUP and press the OK softkey to confirm. Make changes to the primary setup as necessary.
This option allows a specific VTBI and delivery time to be set. The rate necessary to deliver the required volume within the specified time is
calculated and displayed.
1. Stop the infusion. Press
d
2. Select the SET VTBI OVER TIME option using the
3. Adjust the volume to be infused using the
been reached press the OK softkey.
4. Enter the time over which the volume is to be infused using the
5. Press OK softkey to enter the value or BACK to return to the VTBI.
This option allows adjustment of the volume if enabled.
1. Press the
button to access the options menu.
d
2. Select ADJUST ALARM VOLUME.
3. Select HIGH, MEDIUM or LOW using the
4. Press OK softkey to confirm or QUIT to exit screen.
To review pump information:
1. Press the
button to access the options menu.
d
2. Select PUMP DETAILS.
3. Review the information and then press the QUIT softkey.
Basic Features (continued)
Clear Drug Name
keys, press OK to confirm.
f
Setting VTBI over Time
button to access the options menu.
f
f
keys. (Or select BAGS softkey to set the VTBI) When the desired volume has
Adjust Alarm Volume
f
keys.
1000DF00498 Issue 3
keys. Press the OK softkey to confirm the selection.
f
Primary Setup
keys and press the OK softkey.
f
keys . The infusion rate will automatically be calculated.
Pump Details