Electrical Protection
Class I, Type CF (Defibrillation-proof )
Electrical/Mechanical Safety
Complies with IEC/EN60601-1 and IEC/EN60601-2-24.
Electro Magnetic Compatibility (EMC)
Complies with IEC/EN60601-1-2 and IEC/EN60601-2-24.
Electrical Safety
Typical earth leakage current 78µA.
Typical Enclosure Leakage Current (Normal Condition) = 0µA
Typical Protective Earth Resistance = 32mOhms
The above measurements are for guidance only, IEC/EN60601-1
limits are defined below:
Earth Leakage Current (Normal Condition) <= 500µA
Enclosure Leakage Current (Normal Condition) <= 100µA
Protective Earth Resistance <= 200mOhms
Classification - Continuous mode of operation, Portable Equipment
AC Power Supply -
100 - 230 VAC, 50 - 60Hz, 60VA (Maximum).
Fuse Type -
2 X T 1.25 A, slow blowing.
Dimensions -
148mm (w) x 225mm (h) x 148mm (d). Weight: approx. 2.5kg
(excluding power cable).
Protection against fluid ingress -
IPX3 - Protected against spraying water.
Environmental Specifications
Condition
Operating
Temperature
+5°C - +40°C
Humidity
20% - 90%*
Atmospheric
700hPa - 1060 hPa
Pressure
*Non condensing.

Specifications

BATTERY SPECIFICATIONS -
Rechargeable NiMH (Nickel Metal Hydride). Automatically charges
when the pump is connected to AC power.
Battery Life - For a 24 hour battery charge time, the pump at 25ml/h
will have a Mean Time To Battery Empty of 6 hours.
Battery Charging - 2.5 hours to 95%.
Alarm Conditions -
Transport & Storage
-20°C - +50°C
15% - 95%*
500hPa - 1060hPa
Memory Retention -
The electronic memory of the pump will be retained for more
than 2 years with normal use.
1000DF00498 Issue 3
36/42
Warnings
AC POWER FAIL
VTBI DONE
BATTERY LOW
AIR-IN-LINE
TITRATION
SET CLOCK
RATE LOCK
LOG FAILURE
SET SERIAL NUMBER
Prompts
ATTENTION
SET VTBI
SET NOT FITTED
LOCKED
Advisories
DOSE WOULD EXCEED
DOSE UNDER
DOSE NOT PERMITTED
RATE NOT PERMITTED
WEIGHT ABOVE LIMIT
WEIGHT BELOW LIMIT
CONCENTRATION NOT PERMITTED
Alarms
AIR IN LINE (SINGLE BUBBLE)
AIR IN LINE (ACCUMULATED)
DOOR OPEN
DOWNSTREAM OCCLUSION
UPSTREAM OCCLUSION
NO FLOW
FLOW ERROR
FREE FLOW
BATTERY EMPTY
SAFETY CLAMP
SET MISLOAD
FLOW SENSOR
DISCONNECTED
WRONG SET
DOOR CLOSE INCOMPLETE
DO NOT USE
LEVER OPEN