The Alaris® Safety Clamp - CareFusion Alaris GP Guardrails Directions For Use Manual

Volumetric pump
Hide thumbs Also See for Alaris GP Guardrails:
Table of Contents

Advertisement

ALARIS® SAFETY CLAMP**:
Safety Clamp
Tab
Safety Clamp
Slider
SAFETY CLAMP IN OCCLUDED POSITION:
After infusion set is loaded into the pump, opening the door
activates door hooks which will pull the Safety Clamp slider out,
as shown:
NO FLOW
Pushing on the Safety Clamp Slider enables full set flow to the patient. Therefore it is recommended to always close the roller
A
clamp as well.
However, if gravity infusion is required, push up Safety Clamp Tab and push orange Safety Clamp Slider completely into
Frame to enable flow. The gravity infusion can be regulated using the roller clamp on the set.
The Alaris® Safety Clamp
Safety Clamp
Frame
Clamp is in OCCLUDED
POSITION
1000DF00498 Issue 3
SAFETY CLAMP IN NON OCCLUDED POSITION:
When a new infusion set is removed from packaging the Safety
Clamp will be in this position*:
FLOW ENABLED
*
This is necessary to avoid tube damage during storage and to ensure
correct sterilisation and allows immediate priming.
MANUALLY OPERATING THE SAFETY CLAMP
To move the slider into the non occluded position manually,
push up Safety Clamp Tab and push Safety Clamp Slider
completely into Frame:
A
Æ
1 . Push
up
2 . Push
** -
Hereinafter referred to as to as 'Safety Clamp' .
12/42
Clamp in NON
OCCLUDED POSITION