Warnings - CareFusion Alaris GP Guardrails Directions For Use Manual

Volumetric pump
Hide thumbs Also See for Alaris GP Guardrails:
Table of Contents

Advertisement

Display
WRONG SET
DOOR CLOSE
INCOMPLETE
DO NOT USE
LEVER OPEN
Warnings alert the user but may not stop the infusion and are indicated by an audible sound, a flashing amber warning indicator and a message on the
display.
1. Check the display for a warning message. Press
2. Rectify the cause of the warning or proceed with caution.

Warnings:

Display
BATTERY LOW
AC POWER FAIL
VTBI DONE
AIR-IN-LINE
SET CLOCK
TITRATION
RATE LOCK
LOG FAILURE
SET SERIAL NUMBER
*
If pump was on hold the alarm will still be activated but this message will not be displayed.
Alarms (Continued)
Infusion Status Cause
Pump on hold
Safety clamp not detected.
Pump on hold
Safety clamp in non-occluded
position with door open or
obstructed.
Pump on hold / infusion
Internal error has occurred.
stopped
Infusion stopped
Door lever is open
Warnings
to silence the sound for 2 minutes, CANCEL to clear the message.
c
Infusion
Cause
Status
Infusion continues
Less than 30 minutes of battery
life remaining.
Infusion continues*
AC power disconnected or failed. • Reconnect to power supply.
Infusing KVO
Intended VTBI completed.
Air detected in infusion set at
Pump on hold
the start of infusion .
Set not fitted correctly into air in
line detector.
Pump on hold
Date / time not set.
Infusion continues
Rate titration not confirmed.
Infusion continues
Rate lock not confirmed.
Pump on hold
Unable to update event log.
Pump on hold
Serial number not set.
1000DF00498 Issue 3
Action
• Clamp infusion set using roller clamp.
• Check set and close door.
• Replace infusion set. (If necessary)
• Clamp infusion set using roller clamp.
• Investigate and correct set loading.
• Close door.
• Remove pump from use.
• Check door lever.
• Check lever hooks.
• Check lever is not obstructed, if so, free
obstruction.
Action
• Connect to power supply.
• Check power cable.
• Set new VTBI or clear VTBI.
• Ensure set is fitted correctly in the air in line
detector.
• Assess air in infusion set.
• Check fluid level in drip chamber.
• Check level of fluid in container.
• A qualified service engineer must set date /
time.
• Press cancel softkey to continue.
• Confirm or cancel new rate.
• Select YES or NO as required.
• A qualified service engineer may need to
service the
• Contact a qualified service engineer to set the
serial number.
26/42
pump.