Service Configuration Mode - CareFusion Alaris GP Guardrails Directions For Use Manual

Volumetric pump
Hide thumbs Also See for Alaris GP Guardrails:
Table of Contents

Advertisement

This section comprises of a list of options which can be configured. Some can be entered via the pump SERVICE CONFIGURATION menu
(available in Technician Mode) and others through the Guardrails® Editor Software.
Enter the access code on Alaris® GP Guardrails® Volumetric Pump for SERVICE mode, then select SERVICE CONFIGURATION, see the
Technical Service Manual for details.
Use Guardrails® Editor to configure the pump configuration, drug list and units enabled for each data set.
A
Access codes should only be entered by qualified technical personnel .
1. Select DATE & TIME from the SERVICE CONFIGURATION menu using the
2. Press the OK softkey to confirm.
keys to adjust the date displayed, pressing the NEXT softkey to access the next field.
3. Use the
f
4. When the correct date and time are displayed press the OK softkey to return to the SERVICE CONFIGURATION menu.
5. Press the QUIT softkey to return to the SERVICE menu and press
This option is used to add reference text to be shown on the pump start up display.
1. Select PUMP REFERENCE from the SERVICE CONFIGURATION menu using the
keys to enter the text and NEXT to move to the next character.
2. Use the
f
3. When the desired text has been selected press OK softkey to return to the SERVICE CONFIGURATION menu.
4. Press QUIT to exit back to the main SERVICE menu and press
This option is used to set the language of messages shown on the pump display.
1. Select LANGUAGE from the SERVICE CONFIGURATION menu using the
2. Use the
keys to select the language.
f
3. When the desired language has been selected press OK softkey to return to the SERVICE CONFIGURATION menu.
4. Press QUIT to exit back to the main SERVICE menu and press
This option is used to set the backlight and contrast on the pump display.
1. Select BACKLIGHT & CONTRAST from the SERVICE CONFIGURATION menu using the
2. Use the
keys to adjust BACKLIGHT, CONTRAST & DIMMING. The contrast of the display will change when scrolling through
f
the numbers. (Use PARAM to scroll between each option)
3. When the desired value has been reached press the OK softkey, then QUIT to get back to the SERVICE menu and press
power down.
Service Configuration Mode
Date & Time
to exit and power down.
a
Pump Reference Text
to exit and power down.
a
Language
f
to exit and power down.
a
Backlight & Contrast
1000DF00498 Issue 3
22/42
keys and press the OK softkey.
f
keys and press the OK softkey.
f
keys and press the OK softkey.
keys and press the OK softkey.
f
to exit and
a

Advertisement

Table of Contents
loading

Table of Contents