Replacing The Pressure Plate - Leica SM2010 R Instructions For Use Manual

Sliding microtome
Table of Contents

Advertisement

5.4

Replacing the pressure plate

The blade holder SE can be converted from use of narrow blades to use of wide blades, and vice versa.
To do this, both the pressure plate
replaced.
To replace them, proceed as follows:
• Push the knife guard
release the clamp of the pressure plate
• Now carefully pull the insertion aid
(→ Fig. 32-1)
can now be taken off
• To mount another pressure plate, please proceed in the reverse sequence. Only use the pressure
plate together with the matching insertion aid.
Leica SM2010 R
(→ Fig. 32-1)
(→ Fig. 32-3)
towards the right and push the lever
(→ Fig.
1
2
Fig. 32
(→ Fig. 32-2)
(→ Fig.
Fig. 33
and the matching insertion aid
32-1).
3
out to the left
(→ Fig.
33).
Operation
(→ Fig. 32-2)
have to be
(→ Fig. 32-4)
upwards to
4
32). The pressure plate
5
37