SERIES 90 HIGH VOLTAGE SYSTEMS

The Series 90 high voltage test systems use scanner boards that have 64 test points each. These
scanner boards use selectively loaded 96-pin DIN connectors. The connectors are loaded with
pins in the even-numbered outer two rows, A and C. The center row B is not loaded. Therefore,
rows 2, 4, 6, etc., are the only rows with pins. This configuration yields 32 pins per connector.
The first pin of each connector is located in the upper left corner. The second pin is located in
the upper right corner. See Figure 16 for the proper location and orientation of the connectors.
Note:
When making interface cables, be careful to use
these locations and avoid using the pin
numbers stamped on the DIN connectors.
Figure 16. High-voltage Scanner Board Layout
Scanner Board #1
J1
Pin c32
Pin a32
Point 1
Point 2
Pin c2
Pin a2
Point 31
Point 32
J2
Pin c32
Pin a32
Point 33
Point 34
Pin c2
Pin a2
Point 63
Point 64
Series 90HV & H8
(Vertical Board Layout
Page 17
Scanner Board #2
J3
Pin c32
Pin a32
Point 65
Point 66
Pin c2
Pin a2
Point 95
Point 96
J4
Pin c32
Pin a32
Point 97
Point 98
Pin c2
Pin a2
Point 127
Point 128
)