Using Bath Controller With Compatible Brookfield Engineering Software; Using Bath Controller With The Dv-Iii+, Dv-Iii Ultra Or Dv3T Rheometer In Standalone Mode - Ametek BrookField TC Series Operator's Manual

Circulating baths with standard digital temperature controller
Table of Contents

Advertisement

Using Bath Controller with Compatible Brookfield Engineering Software

The controller may communicate with compatible Brookfield Engineering Software. The controller must be
connected to the computer with the appropriate RS232 serial cable.
RS232 Protocol — Rheocalc Software uses the following RS232 protocol:
Data bits — 8
Parity — None
Stop bits — 1
Flow control — None
Baud rate — Selectable (Temperature Controller/PC baud rates must match; 9600 is recommended).
Using Bath Controller with the DV-III+, DV-III Ultra or DV3T Rheometer in Standalone
Mode
The bath can also be controlled directly from the DV-III+, DV-III Ultra or DV3T Rheometer. Use optional
cable (DVP-207) connected from the serial port on the Rheometer to the RS232 serial port on the bath
controller head. Please see additional instructions with your Rheometer Operator's Manual if you wish to
use this mode.
110-512 BEL/EN
2
TX
3
RX
4
DTR
5
GND
6
DSR
7
RTS
8
CTS
9
43