Setting the High Limit Temperature

This sub-menu allows you to limit how high the temperature set point may be set. It also serves as a high
limit safety, alerting you if bath temperature rises above the high limit temperature setting. The High Limit
value may be set from +25° to +175°C / +80° to +350°F; the factory default is 175°C.
To avoid any unwanted alarms or shutdowns during regular operation, the High Limit value should be set
at least 5° higher than the selected control temperature.
To Access: Press the
To Change: Touch the arrows until the desired high limit temperature is displayed.
To Accept: Press
display (approximately 10 seconds).