Adjusting The Temperature Set Point - Ametek BrookField TC Series Operator's Manual

Circulating baths with standard digital temperature controller
Table of Contents

Advertisement

Adjusting the Temperature Set Point

This is the temperature at which the fluid in your Circulating Bath will be maintained. It may be set to one-
tenth of a degree over a range of -50.0° to +170.0°C / -60.0° to +340°F. The factory default set point is
+20.0°C / +68.0°F.
To Change: Press
changes of one degree or more, touch the up/down arrows until the desired set point
temperature is displayed. To make changes of less than a degree, (e.g.,0.50°C), press
a second time. The decimal point will begin flashing. Touch the up/down arrows until
the desired value is displayed.
To Accept: Press
display (approximately 10 seconds).
NOTE: An audible alarm and the words Low Limit or High Limit flashing on the display indicate that the
temperature set point value is outside the Low Limit or High Limit value. The Circulator will continue to
heat/cool until the actual bath temperature reaches the Limit value, at which point operation will stop.
110-512 BEL/EN
. The arrow around the word "SET" will begin flashing. To make
,
, or allow the LCD to return to the main operational
,
Touch and hold up/down
arrows to make large changes
Touch and release arrows or
slide finger up /down scroll bar
to make small changes
Press once to change whole
degree values
Press twice to make changes
of less than a degree (decimal
point will flash)
24

Advertisement

Table of Contents
loading

Table of Contents