Ametek BrookField TC Series Operator's Manual page 46

Circulating baths with standard digital temperature controller
Table of Contents

Advertisement

Legend:
ssss
Set point (multiplied by 10)
tttt
Temperature (multiplied by 10)
mmmm
Ramp time in minutes and tenths of minutes (multiplied by 10)
u
Temperature units digit (F or C)
xxxxxx
Controller ID (TC-501 for all models)
z
Controller status (1 or 2)
Controller Status [z}:
1
Controller ON, in RUN mode
2
Controller OFF, in STANDBY mode
3
No probe connected to controller
5
Temperature reading is above or below allowable limits
6
Temperature input values outside the allowable limits