Fluorescence Light Source (Optional); Uv Shield (Optional; Only Needed For Fluorescence) - Accu-Scope EXI-600 Manual

Inverted research microscope
Hide thumbs Also See for EXI-600:
Table of Contents

Advertisement

Position the turret with the "handle"
facing away from the right side of
the microscope, and slowly slide
the turret onto the track and into
the opening. Ensure that the turret
slides completely into the
microscope.
Close the cover and return the
locking lever into the horizontal
(locked) position.

5.2.11. Fluorescence light source (optional)

Align and insert the fluorescence
light source adapter dovetail into
the receiver on the back of the
microscope (requires epi-
fluorescence vertical illumination is
already installed).
Secure the light source adapter by
tightening the 2mm hex screw
using a 2mm hex screwdriver.

5.2.12. UV shield (optional; only needed for fluorescence)

NOTE: A UV shield is recommended for fluorescence microscopy.
Align the holes in the UV shield
brackets with the tapped holes in
the top of the mainframe and
closest to the observation tube.
Using a 4mm hex screwdriver,
insert and tighten the provided hex
screws to secure the shield in
place.