6.1.10. Programming the LCD panel

After programming, the LCD panel displays the current status of objective position, fluorescence
filter cube turret position, and transmitted light illumination level.
Objective positions: Each of the
six positions on the intelligent
nosepiece can be programmed
into the LCD panel. When in
program mode, choose the
magnification that corresponds to
the objective in that position on the
nosepiece.
Note that the intelligent nosepiece
does not automatically recognize
the presence of an objective.
NOP = No Object Present
Fluorescence filter cube turret
positions: Each of the six positions
on the intelligent filter cube turret
can be programmed into the LCD
panel. When in program mode,
choose the magnification that
corresponds to the filter cube in
that position of the turret.
Arrows on the turret indication the
position as recognized by the LCD
panel. Make a note of these when
the filter cubes are inserted into
the turret.
RECOMMENDATION: Choose an
LCD panel button indicator that
corresponds to the excitation color
of the filter cube being
programmed (e.g., "B" for blue, "G"
for green, "U" for UV or DAPI, "V"
for visible or brightfield, etc.).
Note that the intelligent turret does
not automatically recognize the
presence of a filter cube.
NOP = No Object Present
ACCU-SCOPE
®
73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com
EXI-600 INVERTED MICROSCOPE SERIES
Nosepiece position:
1
Turret position:
1
2
3
4
5
2
3
4
6
5
6
33