Torch Assembly And Disassembly - Agilent Technologies 5800 User Manual

Icp-oes
Hide thumbs Also See for 5800:
Table of Contents

Advertisement

Installation
Parameter
Standard
Fully
Construction
demountable
Injector inner
1.8
diameter (mm)
Injector material
Quartz
Medium with
Outer tube length
slot
Outer tube
Quartz
material
Intermediate tube
Quartz tulip
Recommended
Most sample
application
types
A fully demountable torch can be adapted for another application by replacing the
injector and/or the outer tube with the configuration recommended for that
sample type. Refer to the Agilent Technologies website for ordering information.

Torch Assembly and Disassembly

Both the semi- and fully demountable torch models allow the outer/intermediate
tube-set to be separated from the base/injector for easier cleaning or replacement.
In addition, the fully demountable torch models allow the injector to be removed
from the base for easier cleaning or replacement.
Hot Surface
WARNING
The torch and torch compartment may become extremely hot during
instrument operation and remain hot so for some time after the plasma has
been switched off. Allow the plasma compartment to cool for at least five
minutes before accessing the plasma compartment. Allow the torch to cool
for two minutes before touching the outer tube or attempting to dismantle
the torch.
Agilent 5800 and 5900 ICP-OES User's Guide
Fully Demountable Torches
Inert
High solids
Fully
Fully
demountable
demountable
1.8
2.4
Ceramic
Quartz
Medium with
Medium with
slot
slot
Quartz
Quartz
Quartz tulip
Quartz tulip
Hydrofluoric
Samples with
(HF) acid
high total
digests
dissolved solids
Semi-volatile organics
Fully demountable
1.4
Quartz
Medium with slot
High purity quartz (for
organic solvents)
Quartz tulip
Semi-volatile organic
solvents e.g. ketones,
kerosene, Jet-A1
Volatile organics
Fully demountable
0.8
Quartz
Short (no slot)
High purity quartz (for
organic solvents)
Quartz tulip
Volatile organic
solvents e.g., gasoline
27

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

5900

Table of Contents