Set intensity
After you have selected the desired pre-set, you can set up its
intensity. Pressing the Up-button briefl y will increase the intensity in
small steps of 0,5 mA. When you keep the up-button pressed, the
intensity will increase faster.
Caution: for safety regulations, it is not possible to increase
the intensity beyond 20 mA when keeping the Up-button
pressed. Only when you release the button briefl y can you
increase the intensity again, till the next "hold-point":
20 – 40 – 60 – 70 – 75 – 80 – 85 – 90 – 95 – 100.
Keep to the maximum intensity indicated for the size of the
electrode in Chapter 10.
If there is none or a limited connection between
the device and your skin, and the intensity
comes above 5 mA, a warning sign and a cable
icon will be displayed. The intensity of the
relevant channel will be indicated by "nc". See
"Figure 1 no connection".
When the intensity will be set higher than
7.5 mA, the device disables the channel to
prevent any hazards. See Figure 2 output
disabled.
In this case, check if the cables and the
electrodes are properly connected and if
necessary, replace them. Then retry to set the
intensity again. See the section "Error" for more
information.

Treatment period

Regarding the treatment period, you must strictly adhere to the
instructions from your doctor or pain treating specialist. If you need
a longer treatment period, you should discuss this with your treating
specialist before changing your stimulation patterns.
In consultation with your doctor, you can do exercises or contemporary
activities during the treatment.
i-Pulz 2
Figure 1 no connection
Figure 2 output disabled