10

Errors

Any serious incident that has occurred in relation to the device should
be reported to the manufacturer.

Error-numbers

The device has an internal error detection
system. If an error is detected, the device stops
any active treatment and displays the error
screen. (See Figure 7 error screen.) Table 1 lists
the error numbers and possible solutions.
Error
Device behaviour Cause
1, 2, 3,
Error screen.
4, 5, 6,
All functionalities
9, 18
disabled.
7, 8, 19,
Error screen.
20, 21
All functionalities
disabled.
17
Error screen.
device turns OFF
after 10 sec
10
Error screen
Switch to normal
mode after 10
seconds
11
Error screen
Switch to normal
mode after 10
seconds
12
Error screen
Switch to normal
mode after 10
seconds
26
Internal error
detected
Internal error
detected in the
output signal
Internal error with
battery
Memory error
Memory error
Memory error
Figure 7 error screen
What to do
Turn OFF the device.
When it occurs again by
turning ON, please contact
the distributor.
Turn OFF device and
contact distributor.
Do not use the i-Pulz 2
again until it is checked
by the manufacturer or
distributor.
Turn OFF device and
contact distributor.
Press button, device ready
to use. Settings are set to
default
Press button, device ready
to use. P13, P14, P15 are
set to default
Press button, device ready
to use. Compliance data is
cleared
i-Pulz 2