Table 3. Conducted and Radiated Immunity Guidance
The i-Pulz 2 (T1) is intended for use in the electromagnetic environment
specified below. The customer or the user of i-Pulz 2 (T1) should assure that it
is used in the following environment:
Immunity Test
RF Common
mode/
Conducted
Susceptibility
Radiated RF
Electromagnetic
Field
32
Guidance and Manufacturer's declaration -
electromagnetic emissions
IEC 60601-1-2
Test Level
3 V/m
80 MHz to 2.7
GHz
3 Vrms
150 kHz to 80
MHz
Compliance
level
3 V/m
d =1.2 √P
80 MHz to 800 MHz
d =2.4 √P
800 MHz to 2.5 GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
survey
than the compliance level
in each frequency range
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
Not applicable
Electromagnetic
Environment - guidance
, should be less
a
.
b
i-Pulz 2