Do not remove the patient cable when an output is
active as it could result in an electrical shock. The
i-Pulz 2 checks whether the cable is connected
properly. If not, a warning sign will be displayed.
Display
Table 2 warning or indication screens
28
Device
behaviour
Warning sign +
cable-icon + "nc"
before intensity
value.
Warning sign +
blinking cable-
icon + intensity
value is - - -
and channel is
disabled.
Warning sign +
blinking intensity
value.
The device is
active, but the
backlight is OFF
and it is not
responding to
touch buttons.
Cause
What to do
•
The output is
not connected.
•
The output is
•
disconnected
during
treatment
The output
The device is safe to
signal can't
keep using.
reach the
•
desired
intensity.
•
Key-lock is
Press shortly on the
active
Power button.
Verify if the cable
is connected
to the correct
channel.
Verify if the cable
is connected to
electrodes.
Verify if the
electrodes are
correctly pasted
to the body.
Verify if the
electrodes are
not dried out.
Verify if the skin
is cleansed.
i-Pulz 2