Table 2. Immunity Guidance for Users
The i-Pulz 2 (T1) is intended for use in the electromagnetic environment
specified below. The customer or the user of i-Pulz 2 (T1) should assure that it
is used in the following environment:
Immunity Test
Electrostatic
discharge (ESD)
Electrical fast
transient / burst
Surge
Voltage
dips, short
interruptions,
and voltage
variations on
power supply
input lines
Power
frequency
(50/60 Hz)
magnetic field
i-Pulz 2
Guidance and Manufacturer's declaration -
electromagnetic emissions
IEC 60601-1-2
Test Level
±8 kV contact
±15 kV air
30 A/m
Compliance
level
±8 kV contact
±15 kV air
Not applicable
Not applicable
Not applicable
30 A/m
Electromagnetic
Environment -
guidance
Floors should be wood,
concrete or ceramic
tile. If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
The device is battery
powered
The device is battery
powered
The device is battery
powered
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
31