i-Pulz 2
USER MANUAL

Advertisement

Table of Contents
loading

Summary of Contents for Tens i-Pulz 2

  • Page 1 2 USER MANUAL...
  • Page 2: Table Of Contents

    Read this manual carefully before using the i-Pulz 2 and keep closely to the instructions of your treatment specialist. Table of contents Device information ................4 Intended purpose ................4 Intended patient population ............4 Indications ................4 Contra-indications ..............4 Intended parts of the body .............5 Side effects ..................5...
  • Page 3 Settings menu ................15 Adjust menu .................17 Low and rest intensity .............19 Compliance menu ................19 Controls ..................20 Charging ..................21 Accessories .................21 Cables ..................21 Electrodes available separately ............21 Wireless charger ................21 Optional Wearables ..............22 Contents of package ..............22 Technical information ..............22 Pre-set overview ................24 10 Errors ....................26 Error-numbers................26 Other indications/faults ..............27...
  • Page 4: Device Information

    See the indications. Intended patient population The i-Pulz 2 may be used on all patients with a minimum age of approximately 15 till +99 years indicated for transcutaneous electrical nerve and/or muscle stimulation by a qualified pain treatment specialist.
  • Page 5: Intended Parts Of The Body

    Side effects Electrical stimulation can cause buzzing, tingling, or prickling sensations. See the packaging or leaflet of the electrode for the used materials. Electrodes, gel or patches can cause an allergic reaction, including that of the skin. i-Pulz 2...
  • Page 6: Environment

    • Always check if the i-Pulz 2 is switched off before removing or placing electrodes. • While it is not necessary to store the i-Pulz 2 in its carry case when not being in use or when being transported, it is advisable to do...
  • Page 7: Maintenance

    • If the i-Pulz 2 is to be used on multiple patients, then it is necessary to use new cables and electrodes each time. Professional hygienic maintenance prior to re-use is not required for the i-Pulz 2 device itself.
  • Page 8: Ip Rating

    P13 to P15) and to mask them (render invisible). The masking option is to simplify the use of the device. The i-Pulz 2 will always start with the last used pre-set and, if set, the last used timer-function as well.
  • Page 9: Buttons

    Pressing these buttons allows you to control the intensity. The buttons on the right-hand side of the device control the right-hand side channel and vice versa. These buttons can also be used to set other parameters of the device (see chapter 4). i-Pulz 2...
  • Page 10: The Display

    (see chapter 10). E – Battery indicator This icon indicates the capacity of the battery and is displayed in 5 steps (100%, 75%, 50%, 25%, <5%). F – Intensity display left channel The display shows the set intensity. i-Pulz 2...
  • Page 11: Indication Led

    The display shows the set intensity. H – Graphics section This section is used to clarify the pre-sets used. Every pre-set has a graphic representation of the TENS waves being used. See chapter 4 for device options and other display menus. Indication LED The indication LED lights up during the charging-process or when an error has occurred.
  • Page 12: Cleansing Of The Skin

    Timer button the time increases. By holding the Timer button for 2 seconds, the timer will be disabled. Once the pre-set is started, this timer value will be saved and used automatically when you select that pre-set again. i-Pulz 2...
  • Page 13: Treatment Period

    If you need a longer treatment period, you should discuss this with your treating specialist before changing your stimulation patterns. In consultation with your doctor, you can do exercises or contemporary activities during the treatment. i-Pulz 2...
  • Page 14: Stopping The Treatment

    Stopping the treatment If the timer is activated, the i-Pulz 2 stimulator will switch off both channels automatically when the timer is finished. The treatment can also be stopped manually by holding the pre-set button pressed for 2 seconds.
  • Page 15: Irritation Of The Skin

    Irritation of the skin Despite all the precautions, it is possible that you experience irritation of the skin. If you experience persistent irritation of your skin when using the i-Pulz 2, please contact your treating specialist. Settings menu and advanced possibilities Settings menu...
  • Page 16 By pressing the Timer button, it will switch Month back to the time-overview. When pressing on the Timer button in the time-overview you will be asked if this change should be saved, before the device will go back to the settings menu. i-Pulz 2...
  • Page 17: Adjust Menu

    P1 – P9 or to customize the parameters of the basic pre-sets. In default, these pre-sets are not activated. To activate these pre-sets go the “mask pre-set” menu, so these pre-sets can be unmasked. Configure a user specified pre-set only in consultation with the treating specialist. i-Pulz 2...
  • Page 18 • Use the Timer button to step through the various parameters. • Use the Right Up/Down button to adjust the value of the parameter. By pressing the pre-set button again, the adjust menu will start over again, and you can select the basic pre-sets again. i-Pulz 2...
  • Page 19: Low And Rest Intensity

    (in minutes). The date is based on the time set in the device. If the device-time is inaccurate, it can give a misrepresentation of the date. See settings menu - time to set the correct time in the device. i-Pulz 2...
  • Page 20: Controls

    The i-Pulz 2 makes use of wireless charging technology, which means that no cable needs to be connected to the i-Pulz 2. To get the i-Pulz 2 charged, connect the power adapter to the wireless charge-pad. Plug the power adapter into the mains power socket.
  • Page 21: Accessories

    Wireless charger The i-Pulz 2 makes use of the wireless charging technology of Qi 1.1. This is part of the “ME-SYSTEM” (medical equipment system). The included charge-pad makes use of a double isolated power adapter with 5 Vdc 2000 mA output.
  • Page 22: Optional Wearables

    To make it easier for users to carry the i-Pulz2, optional wearables (a key-cord and a pouch with a belt clip) have been supplied with the device. Contents of package The i-Pulz 2 stimulator is supplied with the following accessories: • i-Pulz 2 nerve stimulator • 2 cables • Wireless charge-pad • Power adapter...
  • Page 23 For more information about electronic and electrical waste equipment disposal, recovery, and collection points, please check your local regulations. This product is in compliance with the Restrictions of the use of certain Hazardous Substances (RoHS). i-Pulz 2...
  • Page 24: Pre-Set Overview

    Pre-set overview i-Pulz 2...
  • Page 25 2...
  • Page 26: Errors

    Turn OFF device and 20, 21 All functionalities detected in the contact distributor. disabled. output signal Do not use the i-Pulz 2 again until it is checked by the manufacturer or distributor. Error screen. Internal error with Turn OFF device and...
  • Page 27: Other Indications/Faults

    The battery is too Put the device on the not turn ON, charger indicator LED is The device does Internal error with Remove the i-Pulz 2 to cool not turn ON, battery down the charger. indicator LED is blinking RED LED Wireless...
  • Page 28 Do not remove the patient cable when an output is active as it could result in an electrical shock. The i-Pulz 2 checks whether the cable is connected properly. If not, a warning sign will be displayed. Display Device Cause...
  • Page 29: Irritated Skin

    This standard defines the rate of immunity for electromagnetic interferences and the maximum electromagnetic radiations for medical equipment. The i-Pulz 2 is manufactured by Rehan Electronics Ltd and adheres to the EN60601-1-2 norm, for both immunity and emission. i-Pulz 2...
  • Page 30 • The i-Pulz 2 contains no parts that can be replaced by the user, and unauthorized repairs can lead to increased emission or reduction of immunity of the device.
  • Page 31 Guidance and Manufacturer’s declaration - electromagnetic emissions The i-Pulz 2 (T1) is intended for use in the electromagnetic environment specified below. The customer or the user of i-Pulz 2 (T1) should assure that it is used in the following environment: Immunity Test...
  • Page 32 Guidance and Manufacturer’s declaration - electromagnetic emissions The i-Pulz 2 (T1) is intended for use in the electromagnetic environment specified below. The customer or the user of i-Pulz 2 (T1) should assure that it is used in the following environment: Immunity Test...
  • Page 33 RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the i-Pulz 2 (T1) is used exceeds the applicable the applicable RF compliance level above, the i-Pulz 2 (T1) should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the i-Pulz 2 (T1).
  • Page 34 Recommended separation distance between portable and mobile RF communications equipment and the i-Pulz 2 (T1) The i-Pulz 2 (T1) is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of...
  • Page 36 Manufacturer Distributor Rehan Electronics Ltd. Aughrim Road Carnew Co-Wicklow Ireland www.rehanelectronics.com Tel: 00353-53-9426742 Fax: 00353-53-9426051...